Pre-sterilisation cleaning of re-usable instruments in general dental practice
- PMID: 17299421
- DOI: 10.1038/bdj.2007.124
Pre-sterilisation cleaning of re-usable instruments in general dental practice
Abstract
Objective: This study examined the policies, procedures, environment and equipment used for the cleaning of dental instruments in general dental practice.
Materials and methods: A total of 179 surgeries were surveyed. This was an observational based study in which the cleaning processes were viewed directly by a trained surveyor. Information relating to surgery policies and equipment was also collected by interview and viewing of records. Data were recorded onto a standardised data collection form prepared for automated reading.
Results: The BDA advice sheet A12 was available in 79% of surgeries visited. The most common method for cleaning dental instruments was manual washing, with or without the use of an ultrasonic bath. Automated washer disinfectors were not used by any surgery visited. The manual wash process was poorly controlled, with 41% of practices using no cleaning agent other than water. Only 2% of surgeries used a detergent formulated for manual washing of instruments. When using ultrasonic baths, the interval that elapsed between changes of the ultrasonic bath cleaning solution ranged from two to 504 hours (median nine hours). Fifty-eight percent of surgeries claimed to have a dedicated area for instrument cleaning, of which 80% were within the patient treatment area. However, in 69% of surgeries the clean and dirty areas were not clearly defined. Virtually all cleaning of dental instruments was undertaken by dental nurses. Training for this was provided mainly by demonstration and observed practice of a colleague. There was little documentation associated with training. Whilst most staff wore gloves when undertaking manual cleaning, 51% of staff did not use eye protection, 57% did not use a mask and 7% used waterproof overalls.
Conclusions: In many dental practices, the cleaning of re-usable dental instruments is undertaken using poorly controlled processes and procedures, which increase the risk of cross infection. Clear and unambiguous advice must be provided to the dental team, especially dental nurses, on appropriate equipment, chemicals and environment for cleaning dental instruments. This should be facilitated by appropriate training programmes and the implementation of quality assurance procedures at each stage of the cleaning process.
Similar articles
-
Sterilization of re-usable instruments in general dental practice.Br Dent J. 2007 Oct 27;203(8):E16. doi: 10.1038/bdj.2007.912. Br Dent J. 2007. PMID: 17906606
-
A method for surveying instrument decontamination procedures in general dental practice.Br Dent J. 2007 Apr 28;202(8):E20; discussion 472-3. doi: 10.1038/bdj.2007.125. Epub 2007 Feb 9. Br Dent J. 2007. PMID: 17299422
-
Cleanability of dental instruments--implications of residual protein and risks from Creutzfeldt-Jakob disease.Br Dent J. 2007 Oct 13;203(7):395-401. doi: 10.1038/bdj.2007.893. Br Dent J. 2007. PMID: 17934424
-
Sterilization. Disciplined microbial control.Dent Clin North Am. 1991 Apr;35(2):339-55. Dent Clin North Am. 1991. PMID: 2032579 Review.
-
The value of ultrasonic cleaners in dental practice.Br Dent J. 1994 Jul 23;177(2):53-6. doi: 10.1038/sj.bdj.4808506. Br Dent J. 1994. PMID: 8060709 Review.
Cited by
-
Infection control in healthcare settings: perspectives for mfDNA analysis in monitoring sanitation procedures.BMC Infect Dis. 2016 Aug 9;16:394. doi: 10.1186/s12879-016-1714-9. BMC Infect Dis. 2016. PMID: 27507586 Free PMC article.
-
Representative survey on the reprocessing of endodontic instruments in Germany.Br Dent J. 2016 May 13;220(9):465-9. doi: 10.1038/sj.bdj.2016.333. Br Dent J. 2016. PMID: 27173705
-
The Effects of Sterilization Procedures on the Cutting Efficiency of Endodontic Instruments: A Systematic Review and Network Meta-Analysis.Materials (Basel). 2021 Mar 22;14(6):1559. doi: 10.3390/ma14061559. Materials (Basel). 2021. PMID: 33810097 Free PMC article. Review.
-
Observance of Sterilization Protocol Guideline Procedures of Critical Instruments for Preventing Iatrogenic Transmission of Creutzfeldt-Jakob Disease in Dental Practice in France, 2017.Int J Environ Res Public Health. 2018 Apr 25;15(5):853. doi: 10.3390/ijerph15050853. Int J Environ Res Public Health. 2018. PMID: 29693615 Free PMC article.
-
Effect of recycling protocol on mechanical strength of used mini-implants.Int J Dent. 2014;2014:424923. doi: 10.1155/2014/424923. Epub 2014 Jul 17. Int J Dent. 2014. PMID: 25136368 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources