Reliability of air displacement plethysmography
- PMID: 17313287
- DOI: 10.1519/00124278-200702000-00030
Reliability of air displacement plethysmography
Abstract
The purpose of this study was to examine the reliability of an air displacement plethysmography device (BOD POD) over trials performed on 3 different days. Subjects consisted of 24 healthy adults (8 men, 16 women), ages 18-38 years, with body weights 46.8-93.6 kg, body mass indexes of 19.1-30.1 kg x m(-2), and percentage body fats (BF) of 7.9-43.1%. Two estimates of BF were performed on 3 days. Paired t-tests revealed no significant within-day differences in body volume (BV), thoracic gas volume (V(TG)), body density (BD), and BF. Correlations between the two V(TG) measures on a day were r = 0.86 for day 1, r = 0.93 for day 2, and r = 0.96 for day 3. BF estimates within a day had high correlations of r = 0.98. Significant differences were found between days for measures of BV, V(TG), BD, and BF. These results indicate a high reliability for within-day estimates of BF and significant differences in between-day estimates of BF using air displacement plethysmography. Reliability of BF may be increased by requiring subjects to practice the procedure for V(TG) measurement.
Similar articles
-
Interdevice variability in percent fat estimates using the BOD POD.Eur J Clin Nutr. 2005 Sep;59(9):996-1001. doi: 10.1038/sj.ejcn.1602202. Eur J Clin Nutr. 2005. PMID: 15970945
-
Within- and between-laboratory precision in the measurement of body volume using air displacement plethysmography and its effect on body composition assessment.Int J Obes Relat Metab Disord. 2004 Jan;28(1):80-90. doi: 10.1038/sj.ijo.0802466. Int J Obes Relat Metab Disord. 2004. PMID: 14710169
-
Body-composition assessment via air-displacement plethysmography in adults and children: a review.Am J Clin Nutr. 2002 Mar;75(3):453-67. doi: 10.1093/ajcn/75.3.453. Am J Clin Nutr. 2002. PMID: 11864850 Review.
-
Body composition by air-displacement plethysmography by using predicted and measured thoracic gas volumes.J Appl Physiol (1985). 1998 Apr;84(4):1475-9. doi: 10.1152/jappl.1998.84.4.1475. J Appl Physiol (1985). 1998. PMID: 9516218 Clinical Trial.
-
Air-displacement plethysmography: here to stay.Curr Opin Clin Nutr Metab Care. 2005 Nov;8(6):624-9. doi: 10.1097/01.mco.0000171127.44525.07. Curr Opin Clin Nutr Metab Care. 2005. PMID: 16205463 Review.
Cited by
-
Physiological and biomechanical responses to exercise on two different types of rowing ergometers in NCAA Division I oarswomen.Eur J Appl Physiol. 2023 Jul;123(7):1529-1541. doi: 10.1007/s00421-023-05172-w. Epub 2023 Mar 16. Eur J Appl Physiol. 2023. PMID: 36929015
-
A Comparison of Body Composition Across Two Phases of the Menstrual Cycle Utilizing Dual-Energy X-Ray Absorptiometry, Air Displacement Plethysmography, and Bioelectrical Impedance Analysis.Int J Exerc Sci. 2017 Dec 1;10(8):1235-1249. doi: 10.70252/HREA6128. eCollection 2017. Int J Exerc Sci. 2017. PMID: 29399250 Free PMC article.
-
A new estimation equation to assess body composition in an athletic population.J Int Soc Sports Nutr. 2025 Dec;22(1):2504578. doi: 10.1080/15502783.2025.2504578. Epub 2025 May 28. J Int Soc Sports Nutr. 2025. PMID: 40434031 Free PMC article.
-
The impact of subject positioning on body composition assessments by air displacement plethysmography evaluated in a heterogeneous sample.PLoS One. 2022 Apr 15;17(4):e0267089. doi: 10.1371/journal.pone.0267089. eCollection 2022. PLoS One. 2022. PMID: 35427395 Free PMC article.
-
Learning Effects in Air Displacement Plethysmography.Life (Basel). 2023 Jun 2;13(6):1315. doi: 10.3390/life13061315. Life (Basel). 2023. PMID: 37374098 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous