. 2007 Apr;118(4):869-76.
doi: 10.1016/j.clinph.2006.12.015. Epub 2007 Feb 20.

Reliability of hearing screening in high-risk neonates: comparative study of otoacoustic emission, automated and conventional auditory brainstem response

Affiliations

Reliability of hearing screening in high-risk neonates: comparative study of otoacoustic emission, automated and conventional auditory brainstem response

A Suppiej et al. Clin Neurophysiol. 2007 Apr.

Abstract

Objective: To compare the diagnostic reliability of automated transient evoked otoacoustic emissions (a-TEOAE), automated auditory brainstem response (a-ABR) and conventional brainstem auditory evoked potential (BAEP/ABR) for identification of hearing loss in high-risk neonates.

Methods: Two hundred and six neonatal intensive care unit (NICU) admitted neonates were tested pre-discharge. Follow-up included a-TEOAE in all children, repetition of a-ABR or BAEP if failed in NICU. Sensitivity and specificity were compared and correlated with auditory risk factors.

Results: BAEP had the highest sensitivity (100%) and specificity (90.8%), a-ABR the lowest (88.9% and 70.6%). A statistically significant difference in risk factors for temporary hearing loss was observed between normal and false positive a-TEOAE and BAEP, but not a-ABR outcome. Differences in specificity between a-ABR and a-TEOAE explain the pattern of "absent a-ABR/present a-TEOAE" in 13.8% of ears.

Conclusions: The BAEP appears the more reliable test for hearing screening of high-risk neonates because of highest sensitivity and specificity and should be used to confirm the diagnosis of "auditory neuropathy" in high-risk neonates. The reliability of a-ABR devices in critically ill neonates needs further investigation.

Significance: This is, to our knowledge, the first attempt to compare the diagnostic reliability of a-TEOAE, a-ABR and BAEP in high-risk neonates.

PubMed Disclaimer

Similar articles

Cited by

Publication types

LinkOut - more resources