[Biomechanical comparison of different stabilisation devices for transforaminal sacral fracture. Is an interlocking device advantageous?]
- PMID: 17318310
- DOI: 10.1007/s00113-007-1236-4
[Biomechanical comparison of different stabilisation devices for transforaminal sacral fracture. Is an interlocking device advantageous?]
Abstract
Background: Reliable osteosynthesis for fractures in the different regions of the human pelvis are described in the literature while there is no common and satisfying treatment for unstable sacral fractures. Because of the posterior pelvic rings special anatomic conditions a local plate osteosynthesis seems to be advantageous. In many fields of modern fracture treatment locking implants show superior results. The prototype of a local locking plate osteosynthesis was compared to a common local plate and two sacroiliac screws.
Methods: The implants were tested using six plastic models of the pelvis and three embalmed human specimens. A Tile C1 fracture was created by disruption of the pubic symphysis and a transforaminal osteotomy. The specimens were exposed to axial loading in an upright single-leg stance with a maximum of 800 N for the plastic models and 200 N for the human specimens. An ultrasonic-based measuring system recorded translations (X, Y, Z) and rotations (alpha, beta, gamma). Parameters such as pattern of motion, translation/rotation, load to failure and remaining dislocation were evaluated.
Results: Concerning most of the evaluated parameters the local plate osteosynthesis was inferior compared with two sacroiliac screws. There were no significant differences between the locking implant and the local plate osteosynthesis. Compared with the two sacroiliac screws the locking implant shows biomechanically equal results but allows greater anterior rotation and remaining dislocation. Because of the lower bone quality, the results from the anatomic specimen tested were not utilisable.
Conclusions: The locking implant is biomechanically an alternative compared with two sacroiliac screws. Problems occurred due to the preset direction of the locking head screws.
Similar articles
-
Biomechanical comparison of three methods of sacral fracture fixation in osteoporotic bone.Spine (Phila Pa 1976). 2010 May 1;35(10):E392-5. doi: 10.1097/BRS.0b013e3181cb4fcd. Spine (Phila Pa 1976). 2010. PMID: 20393387
-
[Biomechanical comparative study of three types of osteosynthesis in the treatment of supra and intercondylar fractures of the humerus in adults].Rev Chir Orthop Reparatrice Appar Mot. 1997;83(3):237-42. Rev Chir Orthop Reparatrice Appar Mot. 1997. PMID: 9255359 French.
-
Biomechanical testing of a concept of posterior pelvic reconstruction in rotationally and vertically unstable fractures.J Bone Joint Surg Br. 2011 Feb;93(2):237-44. doi: 10.1302/0301-620X.93B2.24567. J Bone Joint Surg Br. 2011. PMID: 21282765
-
Principles and clinical application of the locking compression plate (LCP).Acta Chir Orthop Traumatol Cech. 2006 Aug;73(4):221-8. Acta Chir Orthop Traumatol Cech. 2006. PMID: 17026880 Review.
-
A review of locking compression plate biomechanics and their advantages as internal fixators in fracture healing.Clin Biomech (Bristol). 2007 Dec;22(10):1049-62. doi: 10.1016/j.clinbiomech.2007.08.004. Epub 2007 Sep 27. Clin Biomech (Bristol). 2007. PMID: 17904257 Review.
Cited by
-
[Biomechanical principles for treatment of osteoporotic fractures of the pelvis].Unfallchirurg. 2010 Apr;113(4):272-80. doi: 10.1007/s00113-010-1763-2. Unfallchirurg. 2010. PMID: 20309516 Review. German.
-
Biomechanical Comparison of Two Kinds of Internal Fixation in a Type C Zone II Pelvic Fracture Model.Chin Med J (Engl). 2015 Sep 5;128(17):2312-7. doi: 10.4103/0366-6999.163377. Chin Med J (Engl). 2015. PMID: 26315078 Free PMC article.
-
Minimally Invasive Percutaneous TightRope® System Fixation for an Unstable Posterior Pelvic Ring: Clinical Follow-up and Biomechanical Studies.Orthop Surg. 2022 Jun;14(6):1078-1092. doi: 10.1111/os.13261. Epub 2022 Apr 26. Orthop Surg. 2022. PMID: 35470582 Free PMC article. Clinical Trial.
-
Triangular Screw Placement to Treat Dysmorphic Sacral Fragility Fractures in Osteoporotic Bone Results in an Equivalent Stability to Cement-Augmented Sacroiliac Screws-A Biomechanical Cadaver Study.J Clin Med. 2025 Feb 24;14(5):1497. doi: 10.3390/jcm14051497. J Clin Med. 2025. PMID: 40095062 Free PMC article.
-
Management of traumatic spinopelvic dissociations: review of the literature.Eur J Trauma Emerg Surg. 2012 Oct;38(5):517-24. doi: 10.1007/s00068-012-0225-7. Epub 2012 Sep 25. Eur J Trauma Emerg Surg. 2012. PMID: 26816253
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical