A longitudinal study of LASIK flap and stromal thickness with high-speed optical coherence tomography
- PMID: 17320959
- DOI: 10.1016/j.ophtha.2006.09.031
A longitudinal study of LASIK flap and stromal thickness with high-speed optical coherence tomography
Abstract
Objective: To assess corneal anatomic changes after LASIK with a high-speed corneal and anterior segment optical coherence tomography (CAS-OCT) system.
Design: Cross-sectional observational study.
Participants: Fifty-one eyes of 26 healthy persons undergoing LASIK.
Methods: The CAS-OCT prototype operated at a 1.3-mum wavelength and 2000 axial scans/second. The corneas were scanned with a flap profile pattern (horizontal line, 512 axial scans) and a flap map pattern (4 radials, 256 axial scans each). Both patterns are 8 mm long and are centered on the corneal vertex. LASIK flaps were created using either a mechanical microkeratome (Hansatome; Bausch & Lomb, Inc., Rochester, NY) or a femtosecond laser (Pulsion; IntraLase Corp., Irvine, CA). Intraoperative pachymetry was performed using a 50-MHz ultrasound probe. Three OCT scans were obtained on preoperative and post-LASIK visits up to 6 months. An automated algorithm was developed to process the OCT images and to calculate corneal, flap, and stromal bed thickness profiles and maps. The profiles and maps were divided into central (diameter, <2 mm), pericentral (2-5 mm), and transitional (5-7 mm) zones for analysis.
Main outcome measures: Corneal, flap, and stromal bed thicknesses as determined by OCT and ultrasound pachymetry.
Results: The flap interface was best detected in the pericentral zone. One week after surgery, the repeatability of OCT flap and stromal bed thickness measurement was 2 to 7 microm by pooled standard deviation for zones inside a 5-mm diameter. The central flap thickness in 24 Hansatome eyes with a 180-microm setting was 143+/-14 microm by OCT and 131+/-17 microm by ultrasound. In the 8 IntraLase cases with a 120-microm setting, it was 156+/-11 microm by OCT and 160+/-19 microm by ultrasound. Eyes with other settings also were analyzed. There were small systematic changes in flap thickness up to 1 week and bed thickness up to 3 months.
Conclusions: We have developed a method for using high-speed OCT to measure LASIK flap thickness after surgery. The measurement is noncontact, rapid, and repeatable. Profile and map measurements provide more information than point measurements previously demonstrated. This could be valuable for planning LASIK enhancement and characterizing microkeratome performance.
Similar articles
-
Comparison of intraoperative subtraction pachymetry and postoperative anterior segment optical coherence tomography of laser in situ keratomileusis flaps.J Cataract Refract Surg. 2011 Oct;37(10):1879-83. doi: 10.1016/j.jcrs.2011.05.024. Epub 2011 Aug 15. J Cataract Refract Surg. 2011. PMID: 21840682 Clinical Trial.
-
Corneal pachymetry mapping with high-speed optical coherence tomography.Ophthalmology. 2006 May;113(5):792-9.e2. doi: 10.1016/j.ophtha.2006.01.048. Ophthalmology. 2006. PMID: 16650675 Free PMC article.
-
Measurement of LASIK flap thickness with anterior segment optical coherence tomography.J Refract Surg. 2008 Nov;24(9):879-84. doi: 10.3928/1081597X-20081101-05. J Refract Surg. 2008. PMID: 19044227
-
Flap Thickness and the Risk of Complications in Mechanical Microkeratome and Femtosecond Laser In Situ Keratomileusis: A Literature Review and Statistical Analysis.Diagnostics (Basel). 2021 Aug 31;11(9):1588. doi: 10.3390/diagnostics11091588. Diagnostics (Basel). 2021. PMID: 34573930 Free PMC article. Review.
-
Flap and interface complications in LASIK.Curr Opin Ophthalmol. 2002 Aug;13(4):242-5. doi: 10.1097/00055735-200208000-00010. Curr Opin Ophthalmol. 2002. PMID: 12165708 Review.
Cited by
-
Corneal Sublayers Thickness Estimation Obtained by High-Resolution FD-OCT.Int J Biomed Imaging. 2013;2013:989624. doi: 10.1155/2013/989624. Epub 2013 Jun 2. Int J Biomed Imaging. 2013. PMID: 23818895 Free PMC article.
-
Full anterior segment biometry with extended imaging range spectral domain optical coherence tomography at 1340 nm.J Biomed Opt. 2014 Apr;19(4):046013. doi: 10.1117/1.JBO.19.4.046013. J Biomed Opt. 2014. PMID: 24752381 Free PMC article.
-
Clinical applications of anterior segment optical coherence tomography.J Ophthalmol. 2015;2015:605729. doi: 10.1155/2015/605729. Epub 2015 Mar 3. J Ophthalmol. 2015. PMID: 25821589 Free PMC article. Review.
-
Repeatability of layered corneal pachymetry with the artemis very high-frequency digital ultrasound arc-scanner.J Refract Surg. 2010 Sep;26(9):646-59. doi: 10.3928/1081597X-20091105-01. Epub 2009 Nov 16. J Refract Surg. 2010. PMID: 19928698 Free PMC article.
-
Small incision lenticule extraction (SMILE) history, fundamentals of a new refractive surgery technique and clinical outcomes.Eye Vis (Lond). 2014 Oct 16;1:3. doi: 10.1186/s40662-014-0003-1. eCollection 2014. Eye Vis (Lond). 2014. PMID: 26605350 Free PMC article. Review.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources