Comparison of anterior chamber depth measurements using the IOLMaster, scanning peripheral anterior chamber depth analyser, and anterior segment optical coherence tomography
- PMID: 17327264
- PMCID: PMC1954805
- DOI: 10.1136/bjo.2006.113761
Comparison of anterior chamber depth measurements using the IOLMaster, scanning peripheral anterior chamber depth analyser, and anterior segment optical coherence tomography
Abstract
Aim: To compare anterior chamber depth measurements by three non-contact devices-the IOLMaster, scanning peripheral anterior chamber depth analyser (SPAC), and Visante anterior segment optical coherence tomography (AS-OCT) METHODS: Prospective, cross sectional study of 497 phakic subjects over 50 years of age attending a community clinic in Singapore. Anterior chamber depth of the right eye was measured using all three techniques by the same investigator. Depth measurements were made from the corneal epithelium to the anterior lens surface. The values obtained were compared using Bland-Altman analysis.
Results: 232 men and 265 women were examined (mean (SD) age, 63.4 (7.9) years). Mean anterior chamber depth was 3.08 (0.36) mm with IOLMaster, 3.10 (0.44) mm with SPAC, and 3.14 (0.34) mm with AS-OCT. A significant difference was present between the anterior chamber depth measurements recorded by the three devices (p<0.0001). Mean differences between the measurements were: AS-OCT v IOLMaster, 0.062 (0.007) mm (95% limits of agreement, -0.37 to 0.25 mm) (p<0.0001); AS-OCT v SPAC, 0.035 (0.011) mm (-0.44 to 0.51 mm) (p = 0.0001); SPAC v IOLMaster, 0.027 (0.012) mm (-0.57 to 0.50 mm) (p = 0.027).
Conclusions: AS-OCT gave systematically deeper anterior chamber measurements than SPAC and IOL-Master. However, as the differences found were small they are unlikely to be clinically important.
Conflict of interest statement
Conflict of interest: KK has a Japanese patent on SPAC (Japanese patent No 3878164). DSF has been a paid consultant to Carl Zeiss‐Meditec. TA has received research funding and travel support from Carl Zeiss Meditec.
References
-
- Olsen T. Sources of error in intraocular lens power calculation. J Cataract Refract Surg 199218125–129. - PubMed
-
- Naeser K, Boberg‐Ans J, Bargum R. Prediction of pseudo‐phakic anterior chamber depth from pre‐operative data. Acta Ophthalmol 198866433–437. - PubMed
-
- Holladay J T. Standardizing constants for ultrasonic biometry, keratometry, and intraocular lens power calculations. J Cataract Refract Surg 1997231356–1370. - PubMed
-
- Holladay J T. Refractive power calculations for intraocular lenses in the phakic eye. Am J Ophthalmol 199311663–66. - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials