Corrective measures for compromised oxygen delivery during endotracheal tube cuff deflation with high-frequency percussive ventilation
- PMID: 17328825
Corrective measures for compromised oxygen delivery during endotracheal tube cuff deflation with high-frequency percussive ventilation
Abstract
Objective: To determine the effect of endotracheal-tube cuff deflation on airflow and F(IO2) during high-frequency percussive ventilation (HFPV), and explore methods of correcting the cuff-deflation-associated decrease in mean airway pressure and F(IO2) at the carina.
Methods: Using a mechanical lung model in our respiratory research laboratory, we measured circuit pressure near the connection to the endotracheal tube (P(vent)), mean airway pressure (P(aw)), pulsatile tidal volume (V(T)), and F(IO2) at the artificial carina. During cuff deflation we manipulated the pulsatile frequency, pulsatile flow, and the HFPV integral nebulizer. We then assessed 4 methods of correcting the decreased F(IO2) and airway pressure during cuff deflation: (1) oxygen delivery at the inspiratory fail-safe valve, (2) oxygen delivery at the T-piece between the HFPV and the endotracheal tube, (3) continuous activation of the HFPV's integral nebulizer, and (4) oxygen insufflation into the suction channel of the endotracheal tube.
Results: Cuff deflation reduced P(vent), P(aw), pulsatile V(T), and F(IO2). Increasing the pulsatile flow and decreasing the pulsatile frequency further reduced F(IO2) during cuff deflation. Injecting supplemental oxygen at the inspiratory fail-safe valve provided the best F(IO2) increase. Injecting oxygen at the T-piece provided the second best F(IO2) increase. Continuous activation of the integral nebulizer provided the third best F(IO2) increase. Oxygen insufflation to the suction channel was least effective in correcting the F(IO2) decrease caused by cuff deflation.
Conclusion: Cuff-deflation-associated F(IO2), P(aw), and pulsatile V(T) compromise can be partially corrected by any of the 4 methods we studied. Injecting supplemental oxygen at the inspiratory fail-safe valve is the most effective method.
Similar articles
-
Measurement of pulsatile tidal volume, pressure amplitude, and gas flow during high-frequency percussive ventilation, with and without partial cuff deflation.Respir Care. 2007 Jan;52(1):45-9. Respir Care. 2007. PMID: 17194317
-
Effects of Breathing Pattern on Oxygen Delivery Via a Nasal or Pharyngeal Cannula.Respir Care. 2015 Dec;60(12):1804-9. doi: 10.4187/respcare.04173. Epub 2015 Sep 8. Respir Care. 2015. PMID: 26350233
-
An in-vitro study to evaluate high-volume low-pressure endotracheal tube cuff deflation dynamics.Minerva Anestesiol. 2019 Aug;85(8):846-853. doi: 10.23736/S0375-9393.19.13133-1. Epub 2019 Mar 12. Minerva Anestesiol. 2019. PMID: 30871300
-
Right Versus Left Prong Nasal Cannula Flow Delivery and the Effects of Nasal Cycling on Inspired F(IO2) in an Adult Anatomic Model.Respir Care. 2016 Apr;61(4):397-404. doi: 10.4187/respcare.04169. Epub 2015 Nov 17. Respir Care. 2016. PMID: 26577201
-
Inspiratory limb carbon dioxide entrainment during high-frequency oscillatory ventilation: characterization in a mechanical test lung and swine model.Respir Care. 2012 Nov;57(11):1865-72. doi: 10.4187/respcare.01563. Epub 2012 May 14. Respir Care. 2012. PMID: 22613503
Cited by
-
A narrative review of advanced ventilator modes in the pediatric intensive care unit.Transl Pediatr. 2021 Oct;10(10):2700-2719. doi: 10.21037/tp-20-332. Transl Pediatr. 2021. PMID: 34765495 Free PMC article. Review.
-
High-frequency percussive ventilation for airway clearance in cystic fibrosis: a brief report.Lung. 2010 Dec;188(6):511-3. doi: 10.1007/s00408-010-9252-5. Epub 2010 Jul 3. Lung. 2010. PMID: 20602106