Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2007 May-Jun;14(3):295-303.
doi: 10.1197/jamia.M2219. Epub 2007 Feb 28.

Assessing data quality in manual entry of ventilator settings

Affiliations

Assessing data quality in manual entry of ventilator settings

David K Vawdrey et al. J Am Med Inform Assoc. 2007 May-Jun.

Abstract

Objective: To evaluate the data quality of ventilator settings recorded by respiratory therapists using a computer charting application and assess the impact of incorrect data on computerized ventilator management protocols. DESIGN An analysis of 29,054 charting events gathered over 12 months from 678 ventilated patients (1,736 ventilator days) in four intensive care units at a tertiary care hospital.

Measurements: Ten ventilator settings were examined, including fraction of inspired oxygen (Fio (2)), positive end-expiratory pressure (PEEP), tidal volume, respiratory rate, peak inspiratory flow, and pressure support. Respiratory therapists entered values for each setting approximately every two hours using a computer charting application. Manually entered values were compared with data acquired automatically from ventilators using an implementation of the ISO/IEEE 11073 Medical Information Bus (MIB). Data quality was assessed by measuring the percentage of time that the two sources matched. Charting delay, defined as the interval between data observation and data entry, also was measured.

Results: The percentage of time that settings matched ranged from 99.0% (PEEP) to 75.9% (low tidal volume alarm setting). The average charting delay for each charting event was 6.1 minutes, including an average of 1.8 minutes spent entering data in the charting application. In 559 (3.9%) of 14,263 suggestions generated by computerized ventilator management protocols, one or more manually charted setting values did not match the MIB data.

Conclusion: Even at institutions where manual charting of ventilator settings is performed well, automatic data collection can eliminate delays, improve charting efficiency, and reduce errors caused by incorrect data.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Computer charting application in the HELP System used by respiratory therapists to document ventilator settings (far left column) along with other patient-specific ventilator and respiratory data. This screen is specific to the Puritan Bennett 840 ventilator operating in Assist/Control mode. Similar screens were displayed depending on the type of ventilator and mode of operation.
Figure 2
Figure 2
Example of instructions generated by a HELP computerized ventilator management protocol. The protocol logic relied on Fio 2, PEEP, respiratory rate, tidal volume, and pressure support settings that were manually charted by respiratory therapists. Fio 2 = fraction of inspired oxygen, PEEP = positive end-expiratory pressure.
Figure 3
Figure 3
Examples of charting errors that affected the percentage of time that manually charted values were correct. (a) The delayed recording of an Fio 2 setting change is shown near (1), where a setting change occurred at time 14:22, but was not manually charted until 15:25. The Fio 2 setting was correct 86.9% of the time in this 8-hour window. (b) A failure to record an Fio 2 setting change is shown near label (2), and the recording of an incorrect value is shown near (3). The Fio 2 setting was correct 78.1% of the time in this 8-hour window. Fio 2 = fraction of inspired oxygen.
Figure 4
Figure 4
Example of an incorrect Fio 2 setting that was used to generate protocol instructions. Fio 2 = fraction of inspired oxygen.

References

    1. Akhtar SR, Weaver J, Pierson DJ, Rubenfeld GD. Practice variation in respiratory therapy documentation during mechanical ventilation Chest 2003;124:2275-2282. - PubMed
    1. Holt AA, Sibbald WJ, Calvin JE. A survey of charting in critical care units Crit Care Med 1993;21:144-150. - PubMed
    1. Soo Hoo GW, Park L. Variations in the measurement of weaning parameters: a survey of respiratory therapists Chest 2002;121:1947-1955. - PubMed
    1. Dojat M, Harf A, Touchard D, Lemaire F, Brochard L. Clinical evaluation of a computer-controlled pressure support mode Am J Respir Crit Care Med 2000;161:1161-1166. - PMC - PubMed
    1. East TD, Henderson S, Morris AH, Gardner RM. Implementation issues and challenges for computerized clinical protocols for management of mechanical ventilation in ARDS patients Symposium for Computer Applications in Medical Care (SCAMC) 1989;13:583-587.

Publication types

MeSH terms