Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2007 Apr;53(4):729-34.
doi: 10.1373/clinchem.2006.078873. Epub 2007 Mar 1.

Spectrophotometric assay for complex I of the respiratory chain in tissue samples and cultured fibroblasts

Affiliations
Free article

Spectrophotometric assay for complex I of the respiratory chain in tissue samples and cultured fibroblasts

Antoon J M Janssen et al. Clin Chem. 2007 Apr.
Free article

Abstract

Background: A reliable and sensitive complex I assay is an essential tool for the diagnosis of mitochondrial disorders, but current spectrophotometric assays suffer from low sensitivity, low specificity, or both. This deficiency is mainly due to the poor solubility of coenzyme-Q analogs and reaction mixture turbidity caused by the relatively high concentrations of tissue extract that are often required to measure complex I.

Methods: We developed a new spectrophotometric assay to measure complex I in mitochondrial fractions and applied it to muscle and cultured fibroblasts. The method is based on measuring 2,6-dichloroindophenol reduction by electrons accepted from decylubiquinol, reduced after oxidation of NADH by complex I. The assay thus is designed to avoid nonspecific NADH oxidation because electrons produced in these reactions are not accepted by decylubiquinone, resulting in high rotenone sensitivity.

Results: The assay was linear with time and amount of mitochondria. The K(m) values for NADH and 2,6-dichloroindophenol in muscle mitochondria were 0.04 and 0.017 mmol/L, respectively. The highest complex I activities were measured with 0.07 mmol/L decylubiquinone and 3.5 g/L bovine serum albumin. The latter was an essential component of the reaction mixture, increasing the solubility of decylubiquinone and rotenone. In patients with previously diagnosed complex I deficiencies, the new assay detected the complex I deficiencies in both muscle and fibroblasts.

Conclusions: This spectrophotometric assay is reproducible, sensitive, and specific for complex I activity because of its high rotenone sensitivity, and it can be applied successfully to the diagnosis of complex I deficiencies.