Evaluation of phosphenes elicited by extraocular stimulation in normals and by suprachoroidal-transretinal stimulation in patients with retinitis pigmentosa
- PMID: 17342502
- DOI: 10.1007/s00417-007-0563-z
Evaluation of phosphenes elicited by extraocular stimulation in normals and by suprachoroidal-transretinal stimulation in patients with retinitis pigmentosa
Abstract
Background: To determine the efficient parameters to evoke electrical phosphenes is essential for the development of a retinal prosthesis. We studied the efficient parameters in normal subjects and investigated if suprachoroidal-transretinal stimulation (STS) is effective in patients with advanced retinitis pigmentosa (RP) using these efficient parameters.
Methods: The amplitude of pupillary reflex (PR) evoked by transcorneal electrical stimulation (TcES) was determined at different frequencies in eight normal subjects. The relationship between localized phosphenes elicited by transscleral electrical stimulation (TsES) and the pulse parameters was also examined in six normal subjects. The phosphenes evoked by STS were examined in two patients with RP with bare light perception. Biphasic pulses (cathodic first, duration: 0.5 or 1.0 ms, frequency: 20 Hz) were applied through selected channel(s). The size and shape of the phosphenes perceived by the patients were recorded.
Results: The maximum PR was evoked by TcES with a frequency of 20 Hz. The brightest phosphene was elicited by TsES with a pulse train of more than 10 pulses, duration of 0.5-1.0 ms and a frequency of 20 to 50 Hz. In RP patients, localized phosphenes were elicited with a current of 0.3-0.5 mA (0.5 ms) in patient 1 and 0.4 mA (1.0 ms) in patient 2. Two isolated or dumbbell-shaped phosphenes were perceived when the stimulus was delivered through two adjacent channels.
Conclusion: Biphasic pulse trains (> or =10 pulses) with a duration of 0.5-1.0 ms and a frequency of 20-50 Hz were efficient for evoking phosphenes by localized extraocular stimulation in normal subjects. With these parameters, STS is a feasible method to use with a retinal prosthesis even in advanced stages of RPs.
Similar articles
-
Testing of semichronically implanted retinal prosthesis by suprachoroidal-transretinal stimulation in patients with retinitis pigmentosa.Invest Ophthalmol Vis Sci. 2011 Jul 1;52(7):4726-33. doi: 10.1167/iovs.10-6836. Invest Ophthalmol Vis Sci. 2011. PMID: 21436271
-
Electrical stimulation of anterior visual pathways in retinitis pigmentosa.Invest Ophthalmol Vis Sci. 2001 Jan;42(1):291-7. Invest Ophthalmol Vis Sci. 2001. PMID: 11133881
-
Evaluation of residual retinal function by pupillary constrictions and phosphenes using transcorneal electrical stimulation in patients with retinal degeneration.Graefes Arch Clin Exp Ophthalmol. 2006 Oct;244(10):1283-92. doi: 10.1007/s00417-006-0260-3. Epub 2006 Mar 21. Graefes Arch Clin Exp Ophthalmol. 2006. PMID: 16550408
-
[Towards clinical application of a visual prosthesis].Nippon Ganka Gakkai Zasshi. 2009 Mar;113(3):315-42; discussion 343. Nippon Ganka Gakkai Zasshi. 2009. PMID: 19348182 Review. Japanese.
-
Simulating prosthetic vision: I. Visual models of phosphenes.Vision Res. 2009 Jun;49(12):1493-506. doi: 10.1016/j.visres.2009.02.003. Vision Res. 2009. PMID: 19504749 Review.
Cited by
-
Surgical feasibility and biocompatibility of wide-field dual-array suprachoroidal-transretinal stimulation prosthesis in middle-sized animals.Graefes Arch Clin Exp Ophthalmol. 2016 Apr;254(4):661-73. doi: 10.1007/s00417-015-3104-1. Epub 2015 Jul 21. Graefes Arch Clin Exp Ophthalmol. 2016. PMID: 26194404
-
Behavior tests and immunohistochemical retinal response analyses in RCS rats with subretinal implantation of Okayama-University-type retinal prosthesis.J Artif Organs. 2013 Sep;16(3):343-51. doi: 10.1007/s10047-013-0697-1. Epub 2013 Mar 26. J Artif Organs. 2013. PMID: 23529198
-
Electrical Stimulation as a Means for Improving Vision.Am J Pathol. 2016 Nov;186(11):2783-2797. doi: 10.1016/j.ajpath.2016.07.017. Am J Pathol. 2016. PMID: 27643530 Free PMC article. Review.
-
Human Sensation of Transcranial Electric Stimulation.Sci Rep. 2019 Oct 24;9(1):15247. doi: 10.1038/s41598-019-51792-8. Sci Rep. 2019. PMID: 31649289 Free PMC article.
-
Electrical stimulation of cranial nerves in cognition and disease.Brain Stimul. 2020 May-Jun;13(3):717-750. doi: 10.1016/j.brs.2020.02.019. Epub 2020 Feb 23. Brain Stimul. 2020. PMID: 32289703 Free PMC article. Review.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials