Abstract
Fluorescence tomography of diffuse media can yield optimal three-dimensional imaging when multiple projections over 360 degrees geometries are captured, compared with limited projection angle systems such as implementations in the slab geometry. We demonstrate how it is possible to perform noncontact, 360 degrees projection fluorescence tomography of mice using CCD-camera-based detection in free space, i.e., in the absence of matching fluids. This approach achieves high spatial sampling of photons propagating through tissue and yields a superior information content data set compared with fiber-based 360 degrees implementations. Reconstruction feasibility using 36 projections in 10 degrees steps is demonstrated in mice.