Comparative analysis of 5-mm trocars: dilating tip versus non-shielded bladed
- PMID: 17368253
- DOI: 10.1016/j.jmig.2006.09.014
Comparative analysis of 5-mm trocars: dilating tip versus non-shielded bladed
Abstract
Study objective: To examine whether there are statistically significant differences in multiple variables evaluated at different times in the course of surgery and postoperative period when using 5-mm dilating-tip trocars (DTTs) and 5-mm non-shielded-bladed trocars (NSBTs) at randomly selected right or left lateral entry sites on the same patient.
Design: Randomized, single-blinded, controlled trial (Canadian Task Force classification I).
Setting: Center for Women's Care and Reproductive Surgery (CWCRS), Atlanta, Georgia.
Patients: Ninety-four women, median age 45, undergoing laparoscopic surgery at CWCRS for benign gynecologic conditions were randomly assigned to placement of a DTT to the right or left laparoscopic entry site. The NSBT was placed on the contralateral side of the same patient.
Intervention: Each patient had 2 lateral trocars placed, 1 of which was a DTT and the other of which was an NSBT.
Measurements and main results: Comparisons between the sites accessed with the DTT and the NSBT were made by the surgeons at the time of surgery and at the 2-week follow-up, by nurses at 1 and 4 hours after surgery, and by patients at the 2-week follow-up. The nurses and the patients were blinded as to the side of each trocar placement. A visual analog score of 1 to 5 was used for the assessment of 17 studied variables. Questionnaires were standardized and explained to examiners. Wilcoxon's signed-rank test was used for the analysis of time-specific data collected by the same examiner (evaluation by the nurses at 1 and 4 hours after surgery). Friedman's test was applied for analysis of the remaining data. Statistically significant differences were established in ease of placement (chi(2) = 4.691, p = .030) and displacement rate (chi(2) = 7.264, p = .007), in which the NSBT obtained the better results. No statistically significant differences were found in bleeding at the time of placement or removal of the trocars, hematoma/bruising formation, pain, or cosmetic results as assessed by surgeons, nurses, and patients at corresponding stages of intra- and postoperative care.
Conclusions: When used for lateral laparoscopic access in gynecologic surgery, NSBTs were easier to place and had a smaller rate of displacement than DTTs. Despite substantial differences in the design of the trocars, no statistically significant differences in bleeding risk, hematoma/bruising formation, pain, or cosmetic results were established. Individual goals of the surgery and conditions specific to each patient appear to be the best criteria for selection of 1 or the other trocar.
Similar articles
-
Preemptive analgesia installation during gynecologic laparoscopy: a randomized trial.J Minim Invasive Gynecol. 2005 Jul-Aug;12(4):330-5. doi: 10.1016/j.jmig.2005.05.005. J Minim Invasive Gynecol. 2005. PMID: 16036193 Clinical Trial.
-
Minimizing ancillary ports size in gynecologic laparoscopy: a randomized trial.J Minim Invasive Gynecol. 2005 Nov-Dec;12(6):480-5. doi: 10.1016/j.jmig.2005.09.002. J Minim Invasive Gynecol. 2005. PMID: 16337574 Clinical Trial.
-
A comparative, single-blind, randomized trial of pain associated with suction or non-suction drains after gynecologic laparoscopy.J Minim Invasive Gynecol. 2010 Jan-Feb;17(1):16-20. doi: 10.1016/j.jmig.2009.04.010. Epub 2009 May 30. J Minim Invasive Gynecol. 2010. PMID: 19482522 Clinical Trial.
-
Trocar-associated injuries and fatalities: an analysis of 1399 reports to the FDA.J Minim Invasive Gynecol. 2005 Jul-Aug;12(4):302-7. doi: 10.1016/j.jmig.2005.05.008. J Minim Invasive Gynecol. 2005. PMID: 16036187 Review.
-
The role of optical access trocars in laparoscopic surgery.Surg Technol Int. 2005;14:61-7. Surg Technol Int. 2005. PMID: 16525956 Review.
Cited by
-
Blunt versus bladed trocars in laparoscopic surgery: a systematic review and meta-analysis of randomized trials.Surg Endosc. 2013 Jul;27(7):2312-20. doi: 10.1007/s00464-013-2793-y. Epub 2013 Feb 7. Surg Endosc. 2013. PMID: 23389070
-
Trocar types in laparoscopy.Cochrane Database Syst Rev. 2015 Dec 16;2015(12):CD009814. doi: 10.1002/14651858.CD009814.pub2. Cochrane Database Syst Rev. 2015. PMID: 26676093 Free PMC article.
-
Saving time during laparoscopy using a new, wall anchoring trocar device.JSLS. 2010 Apr-Jun;14(2):196-9. doi: 10.4293/108680810X12785289144034. JSLS. 2010. PMID: 20932368 Free PMC article. Clinical Trial.
-
Trocar Injuries in Laparoscopy: Techniques, Tools, and Means for Prevention. A Systematic Review of the Literature.World J Surg. 2016 Oct;40(10):2331-41. doi: 10.1007/s00268-016-3527-9. World J Surg. 2016. PMID: 27146054
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources