Abstract
Objectives: We sought to modify the Serodia HIV-1/HIV-2 particle agglutination assay (PA), a simple and cost-effective HIV assay that is used globally for the detection of HIV antibodies, as a sensitive/less sensitive test (S/LS) to identify recently infected individuals and to estimate HIV incidence.
Methods: The Serodia PA test was modified as an S/LS test (PA-LS) by using HIV antigen-coated gelatin particles at a dilution of 1:68 and a specific diluent, and calibrated using 37 HIV clade B seroconversion panels (309 samples) from Trinidad and from a commercial source that were tested at dilution intervals from 1:10 to 1:80,000. The greatest sensitivity for correctly classifying samples from recent and established infections was determined by receiver operator curve (ROC) analysis.
Results: At a 1:40,000 sample dilution and a days post-seroconversion cutoff of 190 days, the PA-LS test yielded a 97% sensitivity for classifying recent and established infection samples. Furthermore, at a 1:20,000 dilution, the positive predictive value for correctly identifying recently infected individuals was 99%. The PA-LS test offers a 30-44-fold cost saving over currently available S/LS tests.
Conclusion: A modified, low cost and simple-to-perform PA test is appropriate for use in resource-limited countries, and has exhibited excellence in distinguishing recent from established HIV infection.
Similar articles
-
Particle agglutination test "Serodia HIV-1/2" as a novel anti-HIV-1/2 screening test: comparative study on 3311 serum samples.Folia Biol (Praha). 1997;43(4):171-3. Folia Biol (Praha). 1997. PMID: 9338125
-
Anti-HIV-1: extended validation of a modified gelatin particle agglutination assay, and its use with an image analyser.Med Lab Sci. 1991 Oct;48(4):256-60. Med Lab Sci. 1991. PMID: 1811118
-
Modification of rapid human immunodeficiency virus (HIV) antibody assay protocols for detecting recent HIV seroconversion.Clin Diagn Lab Immunol. 2005 Aug;12(8):918-21. doi: 10.1128/CDLI.12.8.918-921.2005. Clin Diagn Lab Immunol. 2005. PMID: 16085908 Free PMC article.
-
[Evolution of IGRA researches].Kekkaku. 2008 Sep;83(9):641-52. Kekkaku. 2008. PMID: 18979999 Review. Japanese.
-
Multispot HIV-1/HIV-2 Rapid Test: advantages over other rapid HIV tests.Expert Rev Mol Diagn. 2007 Sep;7(5):499-505. doi: 10.1586/14737159.7.5.499. Expert Rev Mol Diagn. 2007. PMID: 17892358 Review.
Cited by
-
Diagnosis of Human Immunodeficiency Virus Infection.Clin Microbiol Rev. 2018 Nov 28;32(1):e00064-18. doi: 10.1128/CMR.00064-18. Print 2019 Jan. Clin Microbiol Rev. 2018. PMID: 30487166 Free PMC article. Review.
-
The effect of viral suppression on cross-sectional incidence testing in the johns hopkins hospital emergency department.J Acquir Immune Defic Syndr. 2008 Jun 1;48(2):211-5. doi: 10.1097/QAI.0b013e3181743980. J Acquir Immune Defic Syndr. 2008. PMID: 18520680 Free PMC article.
-
Current and future assays for identifying recent HIV infections at the population level.Med Sci Monit. 2011 May;17(5):RA124-33. doi: 10.12659/msm.881757. Med Sci Monit. 2011. PMID: 21525823 Free PMC article. Review.
-
Moving towards a reliable HIV incidence test - current status, resources available, future directions and challenges ahead.Epidemiol Infect. 2017 Apr;145(5):925-941. doi: 10.1017/S0950268816002910. Epub 2016 Dec 22. Epidemiol Infect. 2017. PMID: 28004622 Free PMC article. Review.
-
Better control of early viral replication is associated with slower rate of elicited antiviral antibodies in the detuned enzyme immunoassay during primary HIV-1C infection.J Acquir Immune Defic Syndr. 2009 Oct 1;52(2):265-72. doi: 10.1097/QAI.0b013e3181ab6ef0. J Acquir Immune Defic Syndr. 2009. PMID: 19525854 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical