Changes in the light intensity of the fiberoptic laryngoscope blade by steam sterilization
- PMID: 17377105
- DOI: 10.1213/01.ane.0000255154.77505.c8
Changes in the light intensity of the fiberoptic laryngoscope blade by steam sterilization
Abstract
Background: I investigated the effects of steam sterilization on light intensity of reusable fiberoptic laryngoscope blades.
Methods: Six new reusable fiberoptic Macintosh laryngoscope blades were steam sterilized. The light intensity was measured, the tip of the blades was photographed, and the light intensity of six new plastic disposable Macintosh laryngoscope blades was also measured.
Results: The light intensity of reusable blades exceeded that of the disposable blades after 20 sterilizations, but was significantly lower after 80 sterilizations. The photographs showed irregular lighting of the blades, which increased with repeated sterilization.
Conclusions: Disposable plastic blades provide more illumination than reusable fiberoptic blades subjected to repeated sterilization.
Similar articles
-
Flexibility and light emission of disposable paediatric Miller 1 laryngoscope blades*.Anaesthesia. 2006 Aug;61(8):792-9. doi: 10.1111/j.1365-2044.2006.04721.x. Anaesthesia. 2006. PMID: 16867093
-
The effect of steam sterilisation at 134 degrees C on light intensity provided by fibrelight Macintosh laryngoscopes.Anaesthesia. 1999 Sep;54(9):875-8. doi: 10.1046/j.1365-2044.1999.00948.x. Anaesthesia. 1999. PMID: 10460560
-
The reliability of laryngoscope lights.Anaesthesia. 2006 May;61(5):488-91. doi: 10.1111/j.1365-2044.2006.04611.x. Anaesthesia. 2006. PMID: 16674626
-
Infection control practices of laryngoscope blades: a review of the literature.AANA J. 2012 Aug;80(4):274-8. AANA J. 2012. PMID: 23251996 Review.
-
Contamination and Disinfection of Rigid Laryngoscopes: A Literature Review.AORN J. 2019 Jul;110(1):49-59. doi: 10.1002/aorn.12724. AORN J. 2019. PMID: 31246299 Review.
Cited by
-
Pattern of anaesthetic equipment contamination and infection prevention in anaesthesia practice at university hospitals.Indian J Anaesth. 2018 Oct;62(10):786-792. doi: 10.4103/ija.IJA_41_18. Indian J Anaesth. 2018. PMID: 30443062 Free PMC article.
-
Effects of battery type and age on performance of rechargeable laryngoscopes.J Anesth. 2013 Oct;27(5):781-4. doi: 10.1007/s00540-013-1618-4. Epub 2013 May 3. J Anesth. 2013. PMID: 23640370
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources