Comparative Study
. 2007 May;118(5):1162-5.
doi: 10.1016/j.clinph.2007.01.016. Epub 2007 Mar 26.

Comparison of standard and pediatric size concentric needle EMG electrodes

Affiliations
Comparative Study

Comparison of standard and pediatric size concentric needle EMG electrodes

Alexander A Brownell et al. Clin Neurophysiol. 2007 May.

Abstract

Objective: To compare motor unit action potential (MUAP) metrics recorded by standard and pediatric size concentric EMG electrodes.

Methods: Commercial electrodes were used to record MUAPs from biceps brachii, first dorsal interosseous and tibialis anterior muscles in normal subjects and those with amyotrophic lateral sclerosis (ALS).

Results: In normal subjects, peak amplitude and area were significantly higher when recorded by the pediatric size electrode in tibialis anterior muscles and peak amplitude recorded in first dorsal interosseous muscles. In ALS subjects, peak amplitude was higher recorded by the pediatric size electrode in tibialis muscle but lower when recorded in first dorsal interosseous muscles.

Conclusions: Differences of MUAP metrics when recording with standard and pediatric size electrodes do not seem to have a clinical relevance.

Significance: Pediatric and standard concentric electrodes record similar MUAP metrics.