Continuous glucose monitoring in interstitial fluid using glucose oxidase-based sensor compared to established blood glucose measurement in rats
- PMID: 17386418
- DOI: 10.1016/j.aca.2006.08.016
Continuous glucose monitoring in interstitial fluid using glucose oxidase-based sensor compared to established blood glucose measurement in rats
Abstract
Glucose monitoring is of importance for success of complex therapeutic interventions in diabetic patients. Its impact on treatment and glycemic control is demonstrated in large clinical trials. Up to eight blood glucose measurements per day are recommended. Notwithstanding, a substantial number of diabetic patients cannot or will not monitor their blood glucose appropriately. Considerable progress in control of disturbed metabolism in diabetic patients can be expected by continuous glucose monitoring. The aim of the study was to evaluate the performance of a new amperometric glucose oxidase-based glucose sensor in vitro and in vivo after subcutaneous implantation into rats. For in vitro testing current output of sensors was measured by exposure to increasing and decreasing glucose concentrations up to 472 mg dL(-1) over a time period of 7 days. After subcutaneous implantation of sensors into interscapular region of male rats glucose in interstitial fluid was evaluated and compared to glucose in arterial blood up to 7 days. Hyper- and hypoglycaemia were induced by intravenous application of glucose and insulin, respectively. Current of each implanted sensor was converted into glucose concentration using the first blood glucose measurement only. A change of current with glucose of 0.35 nA mg(-1)dL(-1) indicates high sensitivity of the sensor in vitro. The response time (90% of steady state) was calculated by approximately 60s. Test strips for blood glucose measurement as reference for sensor readings was found as an appropriate and rapidly available method in rats by comparison with established hexokinase method in an automated lab analyzer with limits of agreement of +32.8 and -25.7 mg dL(-1) in Bland-Altman analysis. In normo- and hypoglycaemic range sensor readings in interstitial fluid correlated well with blood glucose measurements whereas hyperglycaemia was not reflected by the sensor completely when blood glucose was changing rapidly. The data given characterize a sensor with high sensitivity, long term stability and short response time. A single calibration of the sensor is required only in measurement periods up to 7 days. The findings demonstrate that the sensor is a highly promising candidate for assessment in humans.
Similar articles
-
Tissue response to subcutaneous implantation of glucose-oxidase-based glucose sensors in rats.Biosens Bioelectron. 2007 Aug 30;23(1):26-34. doi: 10.1016/j.bios.2007.03.016. Epub 2007 Mar 30. Biosens Bioelectron. 2007. PMID: 17467971
-
Biosensor system for continuous glucose monitoring in fish.Anal Chim Acta. 2009 Feb 2;633(1):90-6. doi: 10.1016/j.aca.2008.11.023. Epub 2008 Nov 18. Anal Chim Acta. 2009. PMID: 19110121
-
Evaluation of subcutaneously-implanted glucose sensors for continuous glucose measurements in hyperglycemic pigs.In Vivo. 2006 Mar-Apr;20(2):195-203. In Vivo. 2006. PMID: 16634519
-
[Subcutaneously implantable glucose sensors in patients with diabetes mellitus; still many problems].Ned Tijdschr Geneeskd. 2002 Jul 13;146(28):1313-6. Ned Tijdschr Geneeskd. 2002. PMID: 12148218 Review. Dutch.
-
Continuous glucose monitoring: an overview of today's technologies and their clinical applications.Int J Clin Pract Suppl. 2002 Jul;(129):75-9. Int J Clin Pract Suppl. 2002. PMID: 12166611 Review.
Cited by
-
Fabrication and Optimization of a Nanoporous Platinum Electrode and a Non-enzymatic Glucose Micro-sensor on Silicon.Sensors (Basel). 2008 Oct 1;8(10):6154-6164. doi: 10.3390/s8106154. Sensors (Basel). 2008. PMID: 27873863 Free PMC article.
-
Continuous Measurement of Lactate Concentration in Human Subjects through Direct Electron Transfer from Enzymes to Microneedle Electrodes.ACS Sens. 2023 Apr 28;8(4):1639-1647. doi: 10.1021/acssensors.2c02780. Epub 2023 Mar 26. ACS Sens. 2023. PMID: 36967522 Free PMC article.
-
3D Printed Microfluidic Device with Integrated Biosensors for Online Analysis of Subcutaneous Human Microdialysate.Anal Chem. 2015 Aug 4;87(15):7763-70. doi: 10.1021/acs.analchem.5b01353. Epub 2015 Jul 20. Anal Chem. 2015. PMID: 26070023 Free PMC article.
-
Simultaneous Monitoring of Sweat and Interstitial Fluid Using a Single Wearable Biosensor Platform.Adv Sci (Weinh). 2018 Aug 2;5(10):1800880. doi: 10.1002/advs.201800880. eCollection 2018 Oct. Adv Sci (Weinh). 2018. PMID: 30356971 Free PMC article.
-
Fully printed and self-compensated bioresorbable electrochemical devices based on galvanic coupling for continuous glucose monitoring.Sci Adv. 2023 Jul 21;9(29):eadi3839. doi: 10.1126/sciadv.adi3839. Epub 2023 Jul 19. Sci Adv. 2023. PMID: 37467335 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous