Systematic error arising from 'Sequential' Standard Addition Calibrations: quantification and correction
- PMID: 17386768
- DOI: 10.1016/j.aca.2007.01.064
Systematic error arising from 'Sequential' Standard Addition Calibrations: quantification and correction
Abstract
Calibrations involving the sequential addition of aliquots of a standard solution to a solution of unknown analyte content may exhibit a systematic error. We show that this systematic error is related to the ratio of the mass fractions in the standard and unknown solutions. This relationship is consistent with experimental results from the determination of lead in aqueous solution by anodic stripping voltammetry using 'Sequential' Standard Addition Calibration (S-SAC). The magnitude of this systematic error has been described mathematically and a correction calculated. These mathematical relationships form the basis for a proposal for best practice in the use of S-SAC.
Similar articles
-
Systematic error arising from 'sequential' standard addition calibrations. 2. Determination of analyte mass fraction in blank solutions.Anal Chim Acta. 2009 Aug 26;648(2):153-6. doi: 10.1016/j.aca.2009.06.061. Epub 2009 Jul 3. Anal Chim Acta. 2009. PMID: 19646577
-
Comparison of quantification strategies for one-point standard addition calibration: the homoscedastic case.Anal Chim Acta. 2012 Feb 24;716:108-11. doi: 10.1016/j.aca.2011.12.041. Epub 2011 Dec 29. Anal Chim Acta. 2012. PMID: 22284884
-
Simple algorithms for nonlinear calibration by the classical and standard additions methods.Analyst. 2005 Mar;130(3):370-8. doi: 10.1039/b411054d. Epub 2005 Jan 28. Analyst. 2005. PMID: 15724167
-
Limits of suspicion, recognition and confirmation as concepts that account for the confirmation transitions at the detection limit for quantification by liquid chromatography-tandem mass spectrometry.J Chromatogr A. 2007 Oct 26;1169(1-2):103-10. doi: 10.1016/j.chroma.2007.08.065. Epub 2007 Sep 2. J Chromatogr A. 2007. PMID: 17880986 Review.
-
Quantitative analyses of cadmium: quality assured?IARC Sci Publ. 1992;(118):73-81. IARC Sci Publ. 1992. PMID: 1303975 Review.
Cited by
-
Traditional Calibration Methods in Atomic Spectrometry and New Calibration Strategies for Inductively Coupled Plasma Mass Spectrometry.Front Chem. 2018 Nov 13;6:504. doi: 10.3389/fchem.2018.00504. eCollection 2018. Front Chem. 2018. PMID: 30483492 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources