Randomized Controlled Trial
. 2007 Jun;143(6):913-919.
doi: 10.1016/j.ajo.2007.02.017. Epub 2007 Apr 2.

Long-term effect of optic edge design in a silicone intraocular lens on posterior capsule opacification

Affiliations
Randomized Controlled Trial

Long-term effect of optic edge design in a silicone intraocular lens on posterior capsule opacification

Wolf Buehl et al. Am J Ophthalmol. 2007 Jun.

Abstract

Purpose: To compare the posterior capsule opacification (PCO) inhibiting effect of the round anterior and sharp posterior optic edge profile of the Clariflex silicone intraocular lens (IOL) [AMO Inc, Santa Ana, California, USA] with that of the double-round edge profile of the SI40 (Phacoflex; AMO Inc, Santa Ana, California, USA) silicone IOL over a period of three years.

Design: Prospective, randomized, double-masked, bilateral clinical trial.

Methods: The study took place at the Department of Ophthalmology, Medical University of Vienna, Austria. Fifty-two patients with age-related cataracts (104 eyes) were included in the study. Each patient received an SI40 IOL (round edges) in one eye and a Clariflex IOL with OptiEdge (sharp posterior optic edge) in the other eye. Follow-up examinations were at one week, one month, six months, and one, two, and three years. Digital retroillumination images were taken of each eye. The amount of PCO was subjectively assessed with the slit-lamp and objectively assessed by automated image analysis software (the computer program Automated Quantification of After-Cataract [AQUA]; Vienna, Austria) one, two, and three years after surgery.

Results: The Clariflex lens showed markedly less PCO at one, two, and three years after surgery. The mean AQUA PCO score was 1.39 for the SI40 and 0.56 for the Clariflex lens after one year, estimated at 1.64 and 0.57 after two years, and at 2.04 and 0.64, respectively, after three years (scale zero to 10; P < .001). The Nd:YAG laser capsulotomy rate was far higher in the SI40 group (Five cases vs one case at three years).

Conclusions: The sharp posterior optic edge profile of the Clariflex silicone IOL led to marked and consistently less PCO than the round-edged SI40 IOL one, two, and three years after surgery.