Axial length measurement using partial coherence interferometry in myopic children: repeatability of the measurement and comparison with refractive components
- PMID: 17401619
- DOI: 10.1007/s10384-006-0410-5
Axial length measurement using partial coherence interferometry in myopic children: repeatability of the measurement and comparison with refractive components
Abstract
Purpose: To evaluate the test-retest repeatability of axial length (AL) measurements made with the IOLMaster in children with a wide range of myopic refractive errors, and to clarify the relationship between the AL and other refractive components.
Methods: The subjects were 95 children (mean age +/- SD, 10.8 +/- 1.3 years; range, 7-13 years; mean refractive error; -4.37 +/- 1.43 D; range, -1.50 to -8.19 D) who participated in a myopia-control trial. The AL of the right eye was measured three times using an IOLMaster, and the mean value was regarded as the representative measurement. After 5 min, the measurement was performed again, and the repeatability was evaluated by analyzing the distribution of differences between the two measurements. The relationships between the AL and age, cycloplegic autorefraction (RE), and corneal radius of curvature (CR) were also examined.
Results: The repeatability of the IOLMaster measurements was +/-0.05 mm (corresponding to a refractive error of +/-0.12 D) and was not affected by age or RE. AL was negatively correlated with RE (RE = -0.68 x AL + 12.74, r = -0.37) and positively correlated with CR (CR = 0.21 x AL + 2.53, r = 0.69). The highest correlation was found between the ratio of AL to CR and RE (AL/CR = -0.04 x RE + 3.08, r = -0.76). No association was observed between age and AL, nor between CR and RE.
Conclusions: The IOLMaster provides high repeatability in AL measurement in myopic children. The associations between AL and each refractive component found in this study were consistent with those in previous studies that used the ultrasound A-mode method, except for slightly higher AL/CR ratios in our study.
Similar articles
-
Comparing ocular biometry and autorefraction measurements from the Myopia Master with the IOLMaster 700 and the Huvitz HRK-8000A autorefractor.Ophthalmic Physiol Opt. 2023 May;43(3):410-417. doi: 10.1111/opo.13101. Epub 2023 Feb 3. Ophthalmic Physiol Opt. 2023. PMID: 36734669
-
Reliability and validity of the partial coherence interferometry for measurement of ocular axial length in children.Eye (Lond). 2006 Sep;20(9):1021-4. doi: 10.1038/sj.eye.6702069. Epub 2005 Aug 12. Eye (Lond). 2006. PMID: 16096655
-
Repeatability of IOLMaster biometry in children.Optom Vis Sci. 2004 Nov;81(11):829-34. doi: 10.1097/01.opx.0000145020.33250.c0. Optom Vis Sci. 2004. PMID: 15545808
-
Reproducibility of optical biometry using partial coherence interferometry : intraobserver and interobserver reliability.J Cataract Refract Surg. 2001 Dec;27(12):1961-8. doi: 10.1016/s0886-3350(01)01214-7. J Cataract Refract Surg. 2001. PMID: 11738911
-
A Comparison of Optical Biometers Used in Children for Myopia Control.Klin Monbl Augenheilkd. 2023 Nov;240(11):1306-1313. doi: 10.1055/a-2117-9335. Epub 2023 Jun 26. Klin Monbl Augenheilkd. 2023. PMID: 37364606 Free PMC article.
Cited by
-
Additive effects of orthokeratology and atropine 0.01% ophthalmic solution in slowing axial elongation in children with myopia: first year results.Jpn J Ophthalmol. 2018 Sep;62(5):544-553. doi: 10.1007/s10384-018-0608-3. Epub 2018 Jul 4. Jpn J Ophthalmol. 2018. PMID: 29974278 Clinical Trial.
-
Effects of scleral cross-linking using genipin on the process of form-deprivation myopia in the guinea pig: a randomized controlled experimental study.BMC Ophthalmol. 2015 Jul 29;15:89. doi: 10.1186/s12886-015-0086-z. BMC Ophthalmol. 2015. PMID: 26220299 Free PMC article.
-
Axial length/corneal radius ratio: association with refractive state and role on myopia detection combined with visual acuity in Chinese schoolchildren.PLoS One. 2015 Feb 18;10(2):e0111766. doi: 10.1371/journal.pone.0111766. eCollection 2015. PLoS One. 2015. PMID: 25693186 Free PMC article.
-
Regulatory roles of differentially expressed MicroRNAs in metabolic processes in negative Lens-induced myopia Guinea pigs.BMC Genomics. 2020 Jan 6;21(1):13. doi: 10.1186/s12864-020-6447-x. BMC Genomics. 2020. PMID: 31906852 Free PMC article.
-
Predictive factors associated with axial length growth and myopia progression in orthokeratology.PLoS One. 2019 Jun 12;14(6):e0218140. doi: 10.1371/journal.pone.0218140. eCollection 2019. PLoS One. 2019. PMID: 31188890 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical