. 1992 Jan;45(1):58-60.
doi: 10.1136/jcp.45.1.58.

Assessment of value of calibrated lyophilised plasmas to determine International Sensitivity Index for coagulometers

Affiliations

Assessment of value of calibrated lyophilised plasmas to determine International Sensitivity Index for coagulometers

K Clarke et al. J Clin Pathol. 1992 Jan.

Abstract

An attempt was made to correct for the effects of coagulometers on the International Sensitivity Index (ISI) in a series of collaborative studies. Modified ISI were derived from the prothrombin time results with coagulometer systems using a range of calibrated plasmas. Two alternative approaches to correction of the ISI were evaluated. The first relied on the consensus orthogonal regression slopes of the prothrombin times for each coagulometer system plotted against the consensus manual results; the second depended on the local individual slope of the prothrombin times for each instrument. The two procedures were compared with the currently recommended method where International Normalised Ratios (INR) are derived from the manual ISI of the thromboplastin. The recommended method gave a significant bias from the manual results with most coagulometers. In contrast, the local correction procedure gave no significant biases, whereas the consensus method did so in a few instances. Both these correction procedures seem more reliable than the recommended method of INR derivation, but the local correction is more accurate and offers a more practical solution by allowing laboratories to determine their own corrected ISI on a range of calibrated plasmas.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Lancet. 1986 Feb 8;1(8476):307-10 - PubMed
    1. Thromb Haemost. 1989 Nov 24;62(3):868-74 - PubMed
    1. J Clin Pathol. 1989 Jan;42(1):97-100 - PubMed
    1. Br J Haematol. 1978 Mar;38(3):391-9 - PubMed
    1. Thromb Haemost. 1990 Jun 28;63(3):424-9 - PubMed

Publication types

LinkOut - more resources