Abstract
Background and purpose: Interfractional prostate motion during radiotherapy due to variation in rectal distension can have negative consequences. The authors investigated the dosimetric consequences of a linear translational position correction based on image guidance when a three-dimensional conformal treatment technique was used.
Material and methods: Planning CTs of seven patients with empty and distended rectum were analyzed. A reference plan for the planning target volume (PTV) and the boost were calculated on the CT dataset with the empty rectum with a standard four-field technique. The treatment plan was transferred to the CT with the distended rectum for an uncorrected setup (referenced to bony anatomy) and a corrected setup after position correction of the isocenter. The dosimetric consequences were analyzed.
Results: Organ motion decreased the coverage of the prostate by the 95% isodose during simulated single treatment fractions by up to -21.0 percentage points (%-p; boost plan) and by up to -14.9%-p for the seminal vesicles (PTV plan). The mean rectum dose increased by up to 18.3%-p (PTV plan). Linear translational correction (mean 6.4 +/- 3.4 mm, maximum 10.8 mm) increased the coverage of the prostate by the 95% isodose by up to 12.7%-p (boost plan), while the mean rectum dose was reduced by up to -8.9%-p (PTV plan). For the complete treatment a reduction of complication probability of the rectum of approximately 5%-p was calculated.
Conclusion: The use of an image guidance system with linear translational correction can improve radiation treatment accuracy for prostate cancer, if geometric changes are within certain limits.
Similar articles
-
Dosimetric consequences of a translational isocenter correction based on image guidance for intensity modulated radiotherapy (IMRT) of the prostate.Phys Med Biol. 2007 Sep 21;52(18):5655-65. doi: 10.1088/0031-9155/52/18/012. Epub 2007 Sep 4. Phys Med Biol. 2007. PMID: 17804887
-
A new strategy for online adaptive prostate radiotherapy based on cone-beam CT.Z Med Phys. 2009;19(4):264-76. doi: 10.1016/j.zemedi.2009.05.007. Epub 2009 Jun 21. Z Med Phys. 2009. PMID: 19962084
-
Online image-guided intensity-modulated radiotherapy for prostate cancer: How much improvement can we expect? A theoretical assessment of clinical benefits and potential dose escalation by improving precision and accuracy of radiation delivery.Int J Radiat Oncol Biol Phys. 2004 Dec 1;60(5):1602-10. doi: 10.1016/j.ijrobp.2004.07.709. Int J Radiat Oncol Biol Phys. 2004. PMID: 15590192
-
[High precision radiotherapy with ultrasonic imaging guidance].Cancer Radiother. 2006 Sep;10(5):231-4. doi: 10.1016/j.canrad.2006.06.005. Epub 2006 Aug 7. Cancer Radiother. 2006. PMID: 16891142 Review. French.
-
Image-guided radiotherapy for localized prostate cancer: treating a moving target.Semin Radiat Oncol. 2008 Jan;18(1):58-66. doi: 10.1016/j.semradonc.2007.09.008. Semin Radiat Oncol. 2008. PMID: 18082589 Review.
Cited by
-
Residual translational and rotational errors after kV X-ray image-guided correction of prostate location using implanted fiducials.Strahlenther Onkol. 2010 Oct;186(10):544-50. doi: 10.1007/s00066-010-2030-8. Epub 2010 Sep 30. Strahlenther Onkol. 2010. PMID: 20936461
-
Intra-fractional uncertainties in image-guided intensity-modulated radiotherapy (IMRT) of prostate cancer.Strahlenther Onkol. 2008 Dec;184(12):668-73. doi: 10.1007/s00066-008-1875-6. Epub 2008 Dec 24. Strahlenther Onkol. 2008. PMID: 19107348
-
Monitoring of tumor oxygenation changes in head-and-neck carcinoma patients breathing a hyperoxic hypercapnic gas mixture with a noninvasive MRI technique.Strahlenther Onkol. 2009 Jan;185(1):19-26. doi: 10.1007/s00066-009-1870-6. Epub 2009 Feb 18. Strahlenther Onkol. 2009. PMID: 19224143
-
Intrafraction motion of the prostate during an IMRT session: a fiducial-based 3D measurement with Cone-beam CT.Radiat Oncol. 2008 Nov 5;3:37. doi: 10.1186/1748-717X-3-37. Radiat Oncol. 2008. PMID: 18986517 Free PMC article.
-
IMRT of prostate cancer: a comparison of fluence optimization with sequential segmentation and direct step-and-shoot optimization.Strahlenther Onkol. 2009 Jun;185(6):379-83. doi: 10.1007/s00066-009-1950-7. Epub 2009 Jun 9. Strahlenther Onkol. 2009. PMID: 19506821 Clinical Trial.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials