Calculation of intratracheal airway pressure in ventilated neonatal piglets with endotracheal tube leaks
- PMID: 17414085
- DOI: 10.1097/01.CCM.0000260244.32756.BB
Calculation of intratracheal airway pressure in ventilated neonatal piglets with endotracheal tube leaks
Abstract
Objective: In ventilated neonates, only the applied pressure of the ventilator is adjusted and monitored. When an endotracheal tube leaks, intratracheal pressure decreases depending on the size of the endotracheal tube and of the leak. Furthermore, an increase in resistance and/or compliance might delay the increase of intratracheal pressure during inspiration and its decline during expiration. Short inspiratory time can cause insufficient ventilation, because intratracheal pressure peak might not be reached. Short expiratory time may lead to air trapping, because intratracheal pressure could not return to baseline. The aim of this study was to develop a mathematical algorithm to calculate intratracheal pressure continuously during ventilation and to evaluate the accuracy of this method.
Design: Prospective, animal study.
Setting: University research laboratory.
Subjects: To verify the mathematical algorithm, eight neonatal piglets (1600-2600 g) were studied under different endotracheal tube leak conditions (45% to 98%). The median compliance and resistance were 1.06 mL/cm H2O/kg and 123 cm H2O/L/sec, respectively.
Interventions: Pressure decreases caused by the different endotracheal tubes were measured in a model while air flow was increased stepwise. Based on these results, a mathematical method was developed to calculate intratracheal pressure under leak conditions continuously in relation to the flow through the endotracheal tube as well as to calculate the values of resistance, compliance, and applied pressure of the ventilator.
Measurements and main results: The intratracheal pressure calculated was compared with the measured intratracheal pressure over time. The differences between measured and calculated intratracheal pressure related to peak applied pressure of the ventilator did not exceed 10%. The medians of absolute amounts of differences between measured and calculated intratracheal pressure were <1 cm H2O.
Conclusions: The accuracy of the calculation of intratracheal pressure ensures adequate monitoring of artificial ventilation, even in the presence of endotracheal tube leaks. This might decrease the risk of barotrauma and improve the effectiveness of ventilation.
Similar articles
-
Correction of compliance and resistance altered by endotracheal tube leaks.Pediatr Crit Care Med. 2003 Jul;4(3):344-52. doi: 10.1097/01.PCC.0000059429.95034.73. Pediatr Crit Care Med. 2003. PMID: 12831418
-
Limitation of measurements of expiratory tidal volume and expiratory compliance under conditions of endotracheal tube leaks.Pediatr Crit Care Med. 2008 Jan;9(1):69-75. doi: 10.1097/01.PCC.0000298660.16328.BA. Pediatr Crit Care Med. 2008. PMID: 18477916
-
Effect of leak around the endotracheal tube on measurements of pulmonary compliance and resistance during mechanical ventilation: a lung model study.Pediatr Pulmonol. 1996 Jul;22(1):35-43. doi: 10.1002/(SICI)1099-0496(199607)22:1<35::AID-PPUL6>3.0.CO;2-J. Pediatr Pulmonol. 1996. PMID: 8856802
-
Special problems in aerosol delivery: artificial airways.Respir Care. 2000 Jun;45(6):636-45. Respir Care. 2000. PMID: 10894456 Review.
-
Anatomy and physiology of tracheostomy.Respir Care. 2005 Apr;50(4):476-82. Respir Care. 2005. PMID: 15807905 Review.