Case Reports
. 2007;34(1):19-22.

Management of air embolism during HeartMate XVE exchange

Affiliations
Case Reports

Management of air embolism during HeartMate XVE exchange

Igor D Gregoric et al. Tex Heart Inst J. 2007.

Abstract

Air embolism is a rare and usually fatal complication of major cardiac surgery. We present a case in which a 45-year-old man supported by a HeartMate(R) XVE left ventricular assist device required a pump exchange due to failure of the device motor. During pump dissection, a massive amount of air entered the systemic circulation. Urgent cannulation for cardiopulmonary bypass was performed, and cardiopulmonary bypass was initiated, followed by profound hypothermia, circulatory arrest, retrograde cerebral perfusion, retrograde coronary sinus perfusion, and then barbiturate coma and steroid therapy. The HeartMate XVE left ventricular assist device was removed, and a HeartMate II was implanted. After 5 days, the patient awoke with left hemiparesis, which nearly resolved with aggressive physical therapy. Forty-four days after the pump exchange operation, the patient was discharged from the hospital with only mild left hemiparesis. Exposure of the left ventricular assist device or its external components requires careful monitoring, because air can enter the pump-particularly in a hypovolemic patient. Rapid response after massive air entry into the left ventricular assist device system, as in our patient, can result in a successful outcome.

PubMed Disclaimer

Figures

None
Fig. 1 After dissection and exposure of the outflow conduit of the HeartMate XVE LVAD, a blood leak was observed at the connection of the conduit to the blood pump (arrow).
None
Fig. 2 The HeartMate XVE outflow conduit shows the position of the leak (arrow). Below the valve, there is a potential for the highest negative pressure; however, above the valve, the pressure is at least as high as the aortic pressure.

References

    1. McCarthy PM, Schmitt SK, Vargo RL, Gordon S, Keys TF, Hobbs RE. Implantable LVAD infections: implications for permanent use of the device. Ann Thorac Surg 1996;61:359–65; discussion 372–3. - PubMed
    1. Holman WL, Pamboukian SV, Blood M, Tallaj JA, McGiffin DC, Kirklin JK. Managing device infections: are we progressing or is infection an insurmountable obstacle? Asaio J 2005;51:452–5. - PubMed
    1. Hammel D, Tjan DT, Scheld HH, Schmid C, Loick M, Deng MC. Successful treatment of a Novacor LVAD malfunction without repeat sternotomy. Thorac Cardiovasc Surg 1998;46: 154–6. - PubMed
    1. McCarthy PM, Smedira NO, Vargo RL, Goormastic M, Hobbs RE, Starling RC, et al. One hundred patients with the HeartMate left ventricular assist device: evolving concepts and technology. J Thorac Cardiovasc Surg 1998;115:904–12. - PubMed
    1. Piccione W Jr. Left ventricular assist device implantation: short and long-term surgical complications. J Heart Lung Transplant 2000;19:S89–94. - PubMed

Publication types