Impact of confidence number on the screening accuracy of the retinomax autorefractor
- PMID: 17435531
- DOI: 10.1097/OPX.0b013e3180339f5a
Impact of confidence number on the screening accuracy of the retinomax autorefractor
Abstract
Purpose: To assess the impact of Retinomax reading confidence number on screening accuracy and to determine whether repeated testing to achieve a higher confidence number improves screening accuracy in preschool children.
Methods: Lay and nurse screeners trained in the use of the Retinomax Autorefractor screened 1452 children enrolled in the Vision in Preschoolers (VIP) Phase II Study. All children also received a comprehensive eye examination. Using statistical comparison of correlated proportions, we compared sensitivity and specificity for detecting any VIP-targeted condition and conditions grouped by severity and by type (amblyopia, strabismus, significant refractive error, and unexplained decreased visual acuity) among three groups of children who had confidence numbers below, at or above the manufacturer's suggested confidence number of 8. The reading with the highest confidence number for each eye was used in the analysis. Each child's confidence number group was defined based on the lower confidence number of the pair of readings for the two eyes. Among the 771 (53.1%) children who had repeated testing either by lay or nurse screeners because of a low confidence number (<8) for one or both eyes in the initial testing, the same analyses were also conducted to compare results between the initial reading with confidence number<8 and repeated test reading with the highest confidence number in the same child. These analyses were based on the failure criteria associated with 90% specificity for detecting any VIP condition in VIP Phase II. We also examined the association between ocular conditions and confidence number. Hochberg procedure was used to adjust the p value for multiple comparisons.
Results: A lower confidence number category was associated with higher sensitivity (0.78, 0.65, and 0.61 for<8, 8, >8, respectively, p=0.04) but much lower specificity (0.64, 0.89, and 0.93, p<0.0001) of detecting any VIP-targeted condition. Through repeated testing, 87% of readings that initially had a confidence number below 8 reached 8 or above, and the increased confidence number that resulted from repeated testing was associated with significantly higher specificity (0.81 vs. 0.86, p=0.002) and a nonsignificant change (by -0.04 to 0.03) in sensitivities. Children with any VIP-targeted condition, significant refractive error, hyperopia, astigmatism, or myopia were more likely to have a low confidence number.
Conclusions: A higher confidence number obtained during Retinomax Autorefractor screening is associated with better screening accuracy. Repeated testing to reach the manufacturer's recommended minimum value is worthwhile in preschool vision screening with the Retinomax. Failure to achieve manufacturer's recommended minimum value through repeated testing should be a factor considered in referring children for a comprehensive eye examination.
Similar articles
-
Vision Screening, Vision Disorders, and Impacts of Hyperopia in Young Children: Outcomes of the Vision in Preschoolers (VIP) and Vision in Preschoolers - Hyperopia in Preschoolers (VIP-HIP) Studies.Asia Pac J Ophthalmol (Phila). 2022 Jan 18;11(1):52-58. doi: 10.1097/APO.0000000000000483. Asia Pac J Ophthalmol (Phila). 2022. PMID: 35044337 Free PMC article. Review.
-
Children unable to perform screening tests in vision in preschoolers study: proportion with ocular conditions and impact on measures of test accuracy.Invest Ophthalmol Vis Sci. 2007 Jan;48(1):83-7. doi: 10.1167/iovs.06-0384. Invest Ophthalmol Vis Sci. 2007. PMID: 17197520 Clinical Trial.
-
Preschool vision screening tests administered by nurse screeners compared with lay screeners in the vision in preschoolers study.Invest Ophthalmol Vis Sci. 2005 Aug;46(8):2639-48. doi: 10.1167/iovs.05-0141. Invest Ophthalmol Vis Sci. 2005. PMID: 16043831 Clinical Trial.
-
Comparison of preschool vision screening tests as administered by licensed eye care professionals in the Vision In Preschoolers Study.Ophthalmology. 2004 Apr;111(4):637-50. doi: 10.1016/j.ophtha.2004.01.022. Ophthalmology. 2004. PMID: 15051194
-
Major review: The underutilization of vision screening (for amblyopia, optical anomalies and strabismus) among preschool age children.Binocul Vis Strabismus Q. 2003;18(4):217-32. Binocul Vis Strabismus Q. 2003. PMID: 14653775 Review.
Cited by
-
Vision Screening, Vision Disorders, and Impacts of Hyperopia in Young Children: Outcomes of the Vision in Preschoolers (VIP) and Vision in Preschoolers - Hyperopia in Preschoolers (VIP-HIP) Studies.Asia Pac J Ophthalmol (Phila). 2022 Jan 18;11(1):52-58. doi: 10.1097/APO.0000000000000483. Asia Pac J Ophthalmol (Phila). 2022. PMID: 35044337 Free PMC article. Review.
-
Astigmatism and myopia in Tohono O'odham Native American children.Optom Vis Sci. 2013 Nov;90(11):1267-73. doi: 10.1097/OPX.0000000000000065. Optom Vis Sci. 2013. PMID: 24100480 Free PMC article.
-
Findings from the Vision in Preschoolers (VIP) Study.Optom Vis Sci. 2009 Jun;86(6):619-23. doi: 10.1097/OPX.0b013e3181a59bf5. Optom Vis Sci. 2009. PMID: 19417714 Free PMC article.
-
Intertester agreement in refractive error measurements.Optom Vis Sci. 2013 Oct;90(10):1128-37. doi: 10.1097/OPX.0000000000000025. Optom Vis Sci. 2013. PMID: 23974664 Free PMC article.
-
Longitudinal change and stability of refractive, keratometric, and internal astigmatism in childhood.Invest Ophthalmol Vis Sci. 2014 Dec 16;56(1):190-8. doi: 10.1167/iovs.14-13898. Invest Ophthalmol Vis Sci. 2014. PMID: 25515577 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical