Tidal volume delivery during high-frequency oscillatory ventilation in adults with acute respiratory distress syndrome
- PMID: 17440422
- DOI: 10.1097/01.CCM.0000266586.04676.55
Tidal volume delivery during high-frequency oscillatory ventilation in adults with acute respiratory distress syndrome
Abstract
Objective: a) Characterize how ventilator and patient variables affect tidal volume during high-frequency oscillatory ventilation; and b) measure tidal volumes in adults with acute respiratory distress syndrome during high-frequency oscillatory ventilation.
Design: Observational study.
Setting: Research laboratory and medical intensive care unit.
Patients: Test lung and patients with acute respiratory distress syndrome.
Interventions: Using a previously validated hot wire anemometer placed in series with a Sensormedics 3100B high-frequency ventilator, an endotracheal tube, and a test lung, tidal volume was measured at different combinations of frequency (4, 6, 8, 10, and 12 Hz), pressure amplitude (50, 60, 70, 80, and 90 cm H2O), mean airway pressure (20, 30, and 40 cm H2O), test lung compliance (10, 30, and 50 mL/cm H2O), endotracheal tube internal diameter (6, 7, and 8 mm), bias flow (20, 30, and 40 L/min), and inspiratory/expiratory ratio (1:2 and 1:1). In patients, tidal volume was measured at baseline ventilator settings and at baseline frequency +/-2 Hz and baseline pressure amplitude +/-10 cm H2O.
Measurements and main results: Measured tidal volumes were 23-225 mL during high-frequency oscillatory ventilation of the test lung. A 2-Hz increase in frequency and a 10-cm H2O increase in pressure amplitude caused a 21.3% +/- 4.1% decrease and 21.4% +/- 3.4% increase in tidal volume, respectively. Decreasing endotracheal tube internal diameter from 8 mm to 7 mm and from 7 mm to 6 mm caused a 15.3% +/- 1.7% and 18.9% +/- 2.1% reduction in tidal volume, respectively. Increasing bias flow from 20 L/min to 30 L/min increased tidal volume by 11.2% +/- 3.9%. Further increases in bias flow, changes in compliance, and changes in mean airway pressure had little effect. Tidal volumes measured in acute respiratory distress syndrome patients were 44-210 mL. A 2-Hz increase in frequency was associated with a 23.1% +/- 6.3% decrease in tidal volume. In contrast to the test lung data, a 10-cm H2O increase in pressure amplitude resulted in only a 5.6% +/- 4.5% increase in tidal volume.
Conclusions: Tidal volumes are not uniformly small during high-frequency oscillatory ventilation. The primary determinant of tidal volume in adults with acute respiratory distress syndrome during high-frequency oscillatory ventilation with the Sensormedics 3100B is frequency. Test lung findings suggest that endotracheal tube internal diameter is also an important determinant of tidal volume.
Comment in
-
High-frequency oscillation: how high should we go?Crit Care Med. 2007 Jun;35(6):1623-4. doi: 10.1097/01.CCM.0000267655.07864.26. Crit Care Med. 2007. PMID: 17522540 No abstract available.
Similar articles
-
Four methods of measuring tidal volume during high-frequency oscillatory ventilation.Crit Care Med. 2006 Mar;34(3):751-7. doi: 10.1097/01.CCM.0000201400.63304.41. Crit Care Med. 2006. PMID: 16505661
-
In vitro performance comparison of the Sensormedics 3100A and B high-frequency oscillatory ventilators.Pediatr Crit Care Med. 2011 Jul;12(4):e176-80. doi: 10.1097/PCC.0b013e3181fe3028. Pediatr Crit Care Med. 2011. PMID: 21037502
-
Feasibility of very high-frequency ventilation in adults with acute respiratory distress syndrome.Crit Care Med. 2008 Apr;36(4):1043-8. doi: 10.1097/01.CCM.0b013e318168fcab. Crit Care Med. 2008. PMID: 18379227
-
A protocol for high-frequency oscillatory ventilation in adults: results from a roundtable discussion.Crit Care Med. 2007 Jul;35(7):1649-54. doi: 10.1097/01.CCM.0000269026.40739.2E. Crit Care Med. 2007. PMID: 17522576
-
High-frequency oscillatory ventilation in adults with acute respiratory distress syndrome.Curr Opin Anaesthesiol. 2012 Feb;25(1):17-23. doi: 10.1097/ACO.0b013e32834ea57b. Curr Opin Anaesthesiol. 2012. PMID: 22157194 Review.
Cited by
-
The Physiological Basis of High-Frequency Oscillatory Ventilation and Current Evidence in Adults and Children: A Narrative Review.Front Physiol. 2022 Apr 26;13:813478. doi: 10.3389/fphys.2022.813478. eCollection 2022. Front Physiol. 2022. PMID: 35557962 Free PMC article. Review.
-
Volumetric xenon-CT imaging of conventional and high-frequency oscillatory ventilation.Acad Radiol. 2009 Jun;16(6):718-25. doi: 10.1016/j.acra.2008.12.003. Epub 2009 Mar 5. Acad Radiol. 2009. PMID: 19268611 Free PMC article.
-
A Novel Non-invasive Predictor Measure of Left Atrial Pressure and Practical Implicates in High Frequency Ventilation: Theory.Cureus. 2024 Jul 6;16(7):e63998. doi: 10.7759/cureus.63998. eCollection 2024 Jul. Cureus. 2024. PMID: 39109117 Free PMC article.
-
Physiologic effects of alveolar recruitment and inspiratory pauses during moderately-high-frequency ventilation delivered by a conventional ventilator in a severe lung injury model.PLoS One. 2017 Sep 29;12(9):e0185769. doi: 10.1371/journal.pone.0185769. eCollection 2017. PLoS One. 2017. PMID: 28961282 Free PMC article.
-
Comparison of high-frequency oscillation and tracheal gas insufflation versus standard high-frequency oscillation at two levels of tracheal pressure.Intensive Care Med. 2010 May;36(5):810-6. doi: 10.1007/s00134-010-1822-8. Epub 2010 Mar 16. Intensive Care Med. 2010. PMID: 20232047 Clinical Trial.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources