Digitization of electrocardiograms by desktop optical scanner
- PMID: 1744545
- DOI: 10.1016/0022-0736(91)90015-e
Digitization of electrocardiograms by desktop optical scanner
Abstract
The fidelity of a semiautomated technique for converting paper electrocardiogram (ECG) tracings to digital form by optical scanning was examined. Sample tracings from one nonmechanical and three mechanical ECG writers (recorders) were used. The optically scanned signals were compared with the digitized version (402 Hz, 12-bit precision) of the original analog signals using time- and frequency-domain correlation coefficients and root mean square error. A total of 261 QRS complexes and 207 RR intervals were examined in 21 leads acquired from 8 patients. When data were low-pass filtered at 25 Hz, the correlation coefficients for the 261 QRS complexes were 0.997 +/- 0.005 (mean +/- SD) for the time domain data, 0.992 +/- 0.010 for the complex frequency domain (amplitude and phase) data, and 0.998 +/- 0.002 for the power spectrum. The corresponding correlations for the 207 RR intervals were 0.993 +/- 0.008, 0.992 +/- 0.008, and 0.993 +/- 0.009. The RMS errors, normalized for signal amplitude, were 2.62 +/- 1.28 (percent +/- SD) for QRS complexes and 1.82 +/- 0.87 for RR intervals. The correlations for the mechanical ECG recorder tracings were the same or better than those of the nonmechanical recorder, and the RMS errors were generally smaller. When data were low-pass filtered at 105 Hz, the correlation coefficients ranged from 0.984 to 0.996 for the QRS complexes and 0.982 to 0.988 for RR intervals. Root mean square errors were 4.54 +/- 2.03 and 2.38 +/- 1.14, respectively. For purposes of arrhythmia analysis by QRS classification, digitization of ECG signals by optical scanning appears equivalent to acquisition via standard analog-to-digital conversion.
Similar articles
-
Image digitization of discontinuous and degraded electrocardiogram paper records using an entropy-based bit plane slicing algorithm.J Electrocardiol. 2018 Jul-Aug;51(4):707-713. doi: 10.1016/j.jelectrocard.2018.05.003. Epub 2018 May 25. J Electrocardiol. 2018. PMID: 29997018
-
Meet the challenge of high-pass filter and ST-segment requirements with a DC-coupled digital electrocardiogram amplifier.J Electrocardiol. 2009 Nov-Dec;42(6):574-9. doi: 10.1016/j.jelectrocard.2009.07.012. Epub 2009 Aug 22. J Electrocardiol. 2009. PMID: 19700169
-
Differences between QT and RR intervals in digital and digitized paper electrocardiograms: contribution of the printer, scanner, and digitization process.J Electrocardiol. 2008 Sep-Oct;41(5):370-5. doi: 10.1016/j.jelectrocard.2008.02.026. Epub 2008 May 14. J Electrocardiol. 2008. PMID: 18482732
-
Comparison of time domain and frequency domain variables from the signal-averaged electrocardiogram: a multivariable analysis.J Am Coll Cardiol. 1988 May;11(5):1041-51. doi: 10.1016/s0735-1097(98)90064-x. J Am Coll Cardiol. 1988. PMID: 3281991 Review.
-
Development of three methods for extracting respiration from the surface ECG: a review.J Electrocardiol. 2014 Nov-Dec;47(6):819-25. doi: 10.1016/j.jelectrocard.2014.07.020. Epub 2014 Aug 4. J Electrocardiol. 2014. PMID: 25194875 Review.
Cited by
-
Novel Tool for Complete Digitization of Paper Electrocardiography Data.IEEE J Transl Eng Health Med. 2013;1:1800107. doi: 10.1109/JTEHM.2013.2262024. IEEE J Transl Eng Health Med. 2013. PMID: 26594601 Free PMC article.