Continuous cardiac output monitoring via arterial pressure waveform analysis following severe hemorrhagic shock in dogs
- PMID: 17452933
- DOI: 10.1097/01.CCM.0000266590.25109.F2
Continuous cardiac output monitoring via arterial pressure waveform analysis following severe hemorrhagic shock in dogs
Abstract
Objectives: To determine agreement and correlation between cardiac output determined by arterial pressure waveform analysis (PulseCO) and the lithium dilution indicator technique (LiDCO) during severe hemorrhagic shock and after fluid resuscitation in dogs.
Design: Prospective experimental study.
Setting: University research laboratory.
Subjects: Twelve adult mongrel dogs.
Interventions: Dogs were anesthetized, and selected arteries and veins were catheterized. Baseline cardiac output was determined by LiDCO and used to calibrate the PulseCO. Hemorrhagic shock was induced by withdrawing blood to achieve and maintain a mean arterial pressure of 30-40 mm Hg for 60 mins, and cardiac output was measured again using both methods. All dogs were resuscitated by administering lactated Ringer's solution intravenously to achieve and maintain a mean arterial pressure between 60 and 70 mm Hg. PulseCO and LiDCO values were measured at 10 and 120 mins after resuscitation.
Measurements and main results: Mean baseline cardiac output was 2.93 +/- 0.45 L/min. PulseCO values overestimated cardiac output compared with LiDCO during hemorrhagic shock (2.25 vs. 0.78 L/min). There were no differences in cardiac output determined by PulseCO and LiDCO at 10 and 120 mins after fluid resuscitation. Bland-Altman analysis suggested that PulseCO values were inaccurate after hemorrhage, producing significant bias with wide limits of agreement and percentage error (1.47 +/- 1.46 L/min; 97%). Bias was small but the limits of agreement and percentage error were large for cardiac output at 10 and 120 mins after resuscitation (-0.1 +/- 1.88 [98%] and -0.17 +/- 1.32 [71%] L/min, respectively). There appeared to be a negative but not significant correlation after hemorrhage (r = -.45; p = .15).
Conclusions: PulseCO determination of cardiac output does not accurately predict rapid decreases in cardiac output or the effects of fluid resuscitation in dogs. Recalibration of PulseCO may be necessary after any apparent or suspected decrease in cardiac preload, afterload, or contractility.
Comment in
-
Pulse contour analysis: fairy tale or new reality?Crit Care Med. 2007 Jul;35(7):1791-2. doi: 10.1097/01.CCM.0000269351.38762.B9. Crit Care Med. 2007. PMID: 17581371 No abstract available.
-
Continuous cardiac output monitoring: really continuous?Crit Care Med. 2007 Oct;35(10):2469-70; author reply 2470. doi: 10.1097/01.CCM.0000284499.49442.6A. Crit Care Med. 2007. PMID: 17885407 No abstract available.
Similar articles
-
Comparison of continuous arterial pressure waveform analysis with the lithium dilution technique to monitor cardiac output in conscious dogs with systemic inflammatory response syndrome.Am J Vet Res. 2009 Nov;70(11):1365-73. doi: 10.2460/ajvr.70.11.1365. Am J Vet Res. 2009. PMID: 19878019
-
Pulse contour cardiac output analysis in a piglet model of severe hemorrhagic shock.Crit Care Med. 2008 Apr;36(4):1189-95. doi: 10.1097/CCM.0B013E31816592A3. Crit Care Med. 2008. PMID: 18401255
-
Comparison of thermodilution, lithium dilution, and pulse contour analysis for the measurement of cardiac output in 3 different hemodynamic states in dogs.J Vet Emerg Crit Care (San Antonio). 2014 Sep-Oct;24(5):562-70. doi: 10.1111/vec.12219. Epub 2014 Aug 20. J Vet Emerg Crit Care (San Antonio). 2014. PMID: 25142925
-
Lithium dilution measurement of cardiac output and arterial pulse waveform analysis: an indicator dilution calibrated beat-by-beat system for continuous estimation of cardiac output.Curr Opin Crit Care. 2002 Jun;8(3):257-61. doi: 10.1097/00075198-200206000-00010. Curr Opin Crit Care. 2002. PMID: 12386506 Review.
-
Resuscitation from severe hemorrhage.Crit Care Med. 1996 Feb;24(2 Suppl):S12-23. Crit Care Med. 1996. PMID: 8608703 Review.
Cited by
-
Continuous and less invasive central hemodynamic monitoring by blood pressure waveform analysis.Am J Physiol Heart Circ Physiol. 2010 Sep;299(3):H584-99. doi: 10.1152/ajpheart.00303.2010. Epub 2010 Jul 9. Am J Physiol Heart Circ Physiol. 2010. PMID: 20622106 Free PMC article. Review.
-
Myths about critical care echocardiography: the ten false beliefs that intensivists should understand.Intensive Care Med. 2015 Jun;41(6):1103-6. doi: 10.1007/s00134-014-3622-z. Epub 2015 Jan 10. Intensive Care Med. 2015. PMID: 25576158 No abstract available.
-
Evaluation of Electrical Cardiometry for Measuring Cardiac Output and Derived Hemodynamic Variables in Comparison with Lithium Dilution in Anesthetized Dogs.Animals (Basel). 2023 Jul 20;13(14):2362. doi: 10.3390/ani13142362. Animals (Basel). 2023. PMID: 37508139 Free PMC article.
-
Uncalibrated pulse power analysis fails to reliably measure cardiac output in patients undergoing coronary artery bypass surgery.Crit Care. 2011;15(1):R76. doi: 10.1186/cc10065. Epub 2011 Feb 28. Crit Care. 2011. PMID: 21356060 Free PMC article.
-
Evaluation of a model-based hemodynamic monitoring method in a porcine study of septic shock.Comput Math Methods Med. 2013;2013:505417. doi: 10.1155/2013/505417. Epub 2013 Mar 25. Comput Math Methods Med. 2013. PMID: 23585774 Free PMC article.