In vivo alteration of Strata valve setting by vagus nerve stimulator-activating magnet
- PMID: 17453775
- DOI: 10.1080/02688690701243217
In vivo alteration of Strata valve setting by vagus nerve stimulator-activating magnet
Abstract
We report the unexpected adjustment of a Strata shunt valve by a vagus nerve stimulator-activating magnet, resulting in symptomatic under-drainage of CSF. Implantable devices using magnets for calibration and activation are becoming more prevalent within neurosurgery and the present case highlights the potential for adverse interactions.
Similar articles
-
Management of a locked Strata valve.J Neurosurg Pediatr. 2009 Apr;3(4):340-3. doi: 10.3171/2008.12.PEDS08306. J Neurosurg Pediatr. 2009. PMID: 19338416
-
Cochlear implantation in the presence of a programmable ventriculoperitoneal shunt.Otol Neurotol. 2009 Sep;30(6):704-7. doi: 10.1097/MAO.0b013e3181b04e3b. Otol Neurotol. 2009. PMID: 19638947
-
Vagus nerve stimulation and magnet use: optimizing benefits.Epilepsy Behav. 2009 Jul;15(3):299-302. doi: 10.1016/j.yebeh.2009.04.002. Epub 2009 May 23. Epilepsy Behav. 2009. PMID: 19361573 Review.
-
Failure of a vagus nerve stimulator following a nearby lightning strike.Pediatr Neurosurg. 2011;47(1):72-3. doi: 10.1159/000327217. Epub 2011 May 5. Pediatr Neurosurg. 2011. PMID: 21546762
-
Intraoperative methods for confirmation of correct placement of the vagus nerve stimulator.Epileptic Disord. 2001 Jun;3(2):75-8. Epileptic Disord. 2001. PMID: 11431169 Review.
Cited by
-
Gastrointestinal damage caused by swallowing multiple magnets.Front Med. 2012 Sep;6(3):280-7. doi: 10.1007/s11684-012-0207-5. Epub 2012 Aug 11. Front Med. 2012. PMID: 22886320 Review.
-
Magnetic toys: forbidden for pediatric patients with certain programmable shunt valves?Childs Nerv Syst. 2009 Feb;25(2):161-4. doi: 10.1007/s00381-008-0770-x. Epub 2008 Dec 5. Childs Nerv Syst. 2009. PMID: 19057906
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources