Patient motion correction for multiplanar, multi-breath-hold cardiac cine MR imaging
- PMID: 17457798
- DOI: 10.1002/jmri.20909
Patient motion correction for multiplanar, multi-breath-hold cardiac cine MR imaging
Abstract
Purpose: To correct for spatial misregistration of multi-breath-hold short-axis (SA), two-chamber (2CH), and four-chamber (4CH) cine cardiac MR (CMR) images caused by respiratory and patient motion.
Materials and methods: Twenty CMR studies from consecutive patients with separate breath-hold 2CH, 4CH, and SA 20-phase cine images were considered. We automatically registered the 2CH, 4CH, and SA images in three dimensions by minimizing the cost function derived from plane intersections for all cine phases. The automatic alignment was compared with manual alignment by two observers.
Results: The processing time for the proposed method was <20 seconds, compared to 14-24 minutes for the manual correction. The initial plane displacement identified by the observers was 2.8 +/- 1.8 mm (maximum = 14 mm). A displacement of >/=5 mm was identified in 15 of 20 studies. The registration accuracy (defined as the difference between the automatic parameters and those obtained by visual registration) was 1.0 +/- 0.9 mm, 1.1 +/- 1.0 mm, 1.1 +/- 1.2 mm, and 2.0 +/- 1.8 mm for 2CH-4CH alignment and SA alignment in the mid, basal, and apical regions, respectively. The algorithm variability was higher in the apex (2.0 +/- 1.9 mm) than in the mid (1.4 +/- 1.4 mm) or basal (1.2 +/- 1.2 mm) regions (ANOVA, P < 0.05).
Conclusion: An automated preprocessing algorithm can reduce spatial misregistration between multiple CMR images acquired at different breath-holds and plane orientations.
(c) 2007 Wiley-Liss, Inc.
Similar articles
-
Comparison of left and right ventricular volume measurement using the Simpson's method and the area length method.Eur J Radiol. 2008 Feb;65(2):270-8. doi: 10.1016/j.ejrad.2007.03.031. Epub 2007 May 10. Eur J Radiol. 2008. PMID: 17498903
-
Automatic 3-D breath-hold related motion correction of dynamic multislice MRI.IEEE Trans Med Imaging. 2010 Mar;29(3):868-78. doi: 10.1109/TMI.2009.2039145. IEEE Trans Med Imaging. 2010. PMID: 20199921
-
Correction of motion artifacts from cardiac cine magnetic resonance images.Acad Radiol. 2005 Oct;12(10):1273-84. doi: 10.1016/j.acra.2005.07.002. Acad Radiol. 2005. PMID: 16179204
-
CMR of ventricular function.Echocardiography. 2007 Feb;24(2):185-93. doi: 10.1111/j.1540-8175.2007.00375.x. Echocardiography. 2007. PMID: 17313554 Review.
-
[Cardiac functional analysis with cine MR imaging].Nihon Rinsho. 1997 Mar;55 Suppl 1:630-3. Nihon Rinsho. 1997. PMID: 9097689 Review. Japanese. No abstract available.
Cited by
-
Highly accelerated cardiac MRI using iterative SENSE reconstruction: initial clinical experience.Int J Cardiovasc Imaging. 2016 Jun;32(6):955-63. doi: 10.1007/s10554-016-0859-3. Epub 2016 Feb 19. Int J Cardiovasc Imaging. 2016. PMID: 26894256
-
Correcting motion in multiplanar cardiac magnetic resonance images.Biomed Eng Online. 2016 Aug 8;15(1):93. doi: 10.1186/s12938-016-0216-8. Biomed Eng Online. 2016. PMID: 27503101 Free PMC article.
-
Cine Cardiac MRI Slice Misalignment Correction Towards Full 3D Left Ventricle Segmentation.Proc SPIE Int Soc Opt Eng. 2018 Feb;10576:1057607. doi: 10.1117/12.2294936. Epub 2018 Mar 12. Proc SPIE Int Soc Opt Eng. 2018. PMID: 30294064 Free PMC article.
-
Quantification of 3D regional myocardial wall thickening from gated magnetic resonance images.J Magn Reson Imaging. 2010 Feb;31(2):317-27. doi: 10.1002/jmri.22033. J Magn Reson Imaging. 2010. PMID: 20099344 Free PMC article.
-
Free-breathing 2D radial cine MRI with respiratory auto-calibrated motion correction (RAMCO).Magn Reson Med. 2023 Mar;89(3):977-989. doi: 10.1002/mrm.29499. Epub 2022 Nov 8. Magn Reson Med. 2023. PMID: 36346081 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous