. 2007 Jul;33(7):1139-1147.
doi: 10.1007/s00134-007-0630-2. Epub 2007 Apr 26.

Sleep during proportional-assist ventilation with load-adjustable gain factors in critically ill patients

Affiliations

Sleep during proportional-assist ventilation with load-adjustable gain factors in critically ill patients

C Alexopoulou et al. Intensive Care Med. 2007 Jul.

Abstract

Background: Proportional-assist ventilation with load-adjustable gain factors (PAV+) automatically adjusts the flow and volume assist to represent constant fractions of resistance and elastance of the respiratory system, respectively. Resistance and elastance are calculated at random intervals of 4-10 breaths, by applying a 300 ms pause maneuver at the end of selected inspirations.

Objectives: To determine whether the large number of end-inspiratory occlusions during PAV+ operation influences sleep quality in critically ill patients who exhibited good patient-ventilator synchrony during pressure support (PS, baseline).

Methods: One and two nights' polysomnography was performed in sedated (protocol A, n=11) and non-sedated (protocol B, n=9) patients, respectively, while respiratory variables were continuously recorded. In each protocol the patients were ventilated with PAV+ and PS at two levels of assist (baseline and high).

Results: In both protocols sleep quality did not differ between the modes of support or the assist levels. In sedated patients sleep efficiency was slightly but significantly higher with PAV+ than with high PS, while it did not differ between modes in non-sedated patients. The two modes of support had comparable effects on respiratory variables. Independent of the mode of support and particularly at high assist, a significant proportion of patients developed periodic breathing during sleep (27% in protocol A and 44% in protocol B).

Conclusion: In patients exhibiting good patient-ventilator synchrony during PS, the large number of short-term end-inspiratory occlusions with PAV+ operation did not adversely influence sleep quality. With both modes high assist may cause unstable breathing during sleep.

PubMed Disclaimer

References

    1. Am J Respir Crit Care Med. 2000 Mar;161(3 Pt 1):819-26 - PubMed
    1. Am J Respir Crit Care Med. 2001 Mar;163(4):829-39 - PubMed
    1. Clin Neurophysiol. 2002 Nov;113(11):1798-805 - PubMed
    1. Intensive Care Med. 2003 Jul;29(7):1126-33 - PubMed
    1. Intensive Care Med. 2006 May;32(5):692-9 - PubMed

LinkOut - more resources