Central and peripheral pachymetry measurements according to age using the Pentacam rotating Scheimpflug camera
- PMID: 17466857
- DOI: 10.1016/j.jcrs.2006.12.025
Central and peripheral pachymetry measurements according to age using the Pentacam rotating Scheimpflug camera
Abstract
Purpose: To investigate the mean values and standard deviations as well as the reliability of consecutive examinations of central and peripheral corneal thickness measurements according to age using the Pentacam rotating Scheimpflug camera (Oculus, Inc.).
Setting: Department of Ophthalmology, University of Heidelberg, Heidelberg, Germany.
Methods: Seventy-six healthy volunteers were enrolled in a clinical prospective study. Three consecutive Pentacam measurements of 1 eye per subject were taken. Evaluated were the central corneal thickness (CCT), corneal thickness at 4 peripheral points (3.0 mm superior, inferior, nasal, and temporal), and the thinnest point of the cornea. The volunteers were then assigned to 3 groups to assess the influence of increasing age on the study parameters.
Results: The mean age of the subjects was 46.6 years +/- 16.8 (SD). The mean CCT was 539.62 +/- 31.87 microm. Peripherally, the corneal thickness was between 11% and 19% higher than centrally, with the superior cornea being the thickest followed by the nasal, the inferior, and the temporal cornea. The thinnest point was located in the inferotemporal quadrant in 92% of eyes and in the superotemporal quadrant in 8%. There was no correlation between age and corneal thickness. Minor mean standard deviations of consecutive measurements were noted in the corneal center (4.33 microm), increasing significantly toward the periphery (mean 8.31 microm). Increasing age was not associated with decreasing reliability.
Conclusions: Using the Pentacam, it was possible to acquire information about corneal thickness across the entire cornea. No correlation was found between increasing age and change in peripheral or CCT. Good reliability was noted for pachymetry measurements decreasing slightly toward the periphery, which was independent of age.
Comment in
-
Reliability of peripheral corneal pachymetry with the Oculus Pentacam.J Cataract Refract Surg. 2008 Jan;34(1):7; author reply 8. doi: 10.1016/j.jcrs.2007.08.042. J Cataract Refract Surg. 2008. PMID: 18165061 No abstract available.
Similar articles
-
[Central and peripheral corneal pachymetry--standard evaluation with the Pentacam system].Klin Monbl Augenheilkd. 2005 Feb;222(2):117-22. doi: 10.1055/s-2005-857908. Klin Monbl Augenheilkd. 2005. PMID: 15719315 Clinical Trial. German.
-
Central and peripheral corneal thickness measured with optical coherence tomography, Scheimpflug imaging, and ultrasound pachymetry in normal, keratoconus-suspect, and post-laser in situ keratomileusis eyes.J Cataract Refract Surg. 2009 Jun;35(6):1055-62. doi: 10.1016/j.jcrs.2009.01.022. J Cataract Refract Surg. 2009. PMID: 19465292
-
Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Scheimpflug camera and ultrasound pachymetry.Am J Ophthalmol. 2007 Nov;144(5):712-718. doi: 10.1016/j.ajo.2007.07.021. Epub 2007 Sep 17. Am J Ophthalmol. 2007. PMID: 17869208
-
Measuring the cornea: the latest developments in corneal topography.Curr Opin Ophthalmol. 2007 Jul;18(4):325-33. doi: 10.1097/ICU.0b013e3281ca7121. Curr Opin Ophthalmol. 2007. PMID: 17568210 Review.
-
[Pentacam--the new way for anterior eye segment imaging and mapping].Klin Oczna. 2005;107(10-12):728-31. Klin Oczna. 2005. PMID: 16619832 Review. Polish.
Cited by
-
Comparison of corneal thickness in patients with dry eye disease using the Pentacam rotating Scheimpflug camera and anterior segment optical coherence tomography.PLoS One. 2020 Feb 3;15(2):e0228567. doi: 10.1371/journal.pone.0228567. eCollection 2020. PLoS One. 2020. PMID: 32012201 Free PMC article.
-
Comparison of Mean Corneal Power of Annular Rings and Zones Using Swept-Source Optical Coherence Tomography.Diagnostics (Basel). 2022 Mar 19;12(3):754. doi: 10.3390/diagnostics12030754. Diagnostics (Basel). 2022. PMID: 35328307 Free PMC article.
-
[Scheimpflug and topography systems in ophthalmologic diagnostics].Ophthalmologe. 2008 Sep;105(9):810-7. doi: 10.1007/s00347-008-1821-0. Ophthalmologe. 2008. PMID: 18751984 German.
-
The precision and agreement of corneal thickness and keratometry measurements with SS-OCT versus Scheimpflug imaging.Eye Vis (Lond). 2020 Jun 9;7:32. doi: 10.1186/s40662-020-00197-0. eCollection 2020. Eye Vis (Lond). 2020. PMID: 32528997 Free PMC article.
-
Diagnostic Techniques to Increase the Safety of Phakic Intraocular Lenses.Diagnostics (Basel). 2023 Jul 27;13(15):2503. doi: 10.3390/diagnostics13152503. Diagnostics (Basel). 2023. PMID: 37568866 Free PMC article. Review.