Repeatability and reproducibility of anterior chamber angle measurement with anterior segment optical coherence tomography
- PMID: 17475709
- PMCID: PMC2095411
- DOI: 10.1136/bjo.2007.118901
Repeatability and reproducibility of anterior chamber angle measurement with anterior segment optical coherence tomography
Abstract
Aim: To evaluate the repeatability and reproducibility of anterior chamber angle measurement obtained by anterior segment optical coherence tomography.
Methods: Twenty-five normal subjects were invited for anterior chamber angle imaging with an anterior segment optical coherence tomography (OCT) on one randomly selected eye in three separate visits within a week. Each eye was imaged three times under room light (light intensity = 368 lux) and three times in the dark during the first visit. In the subsequent visits, each eye was imaged once in the light and once in the dark. The angle opening distance (AOD 500) and the trabecular-iris angle (TIA 500) were measured by a single observer. Only the nasal angle was analysed. Intrasession and intersession within-subject standard deviation (Sw), precision (1.96xSw), coefficient of variation (CVw) (100xSw/overall mean), and intraclass correlation coefficient (ICC) were calculated to evaluate repeatability and reproducibility.
Results: For intrasession repeatability, the Sw, precision, CVw and ICC of AOD/TIA were 45 microm/2.4 degrees , 88 microm/4.7 degrees , 5.8%/4.8% and 0.97/0.95 in the light; and 45 microm/2.1 degrees , 88 microm/4.2 degrees , 7.0%/5.0% and 0.98/0.97 in the dark. For intersession reproducibility, the Sw, precision, CVw and ICC of AOD/TIA were 79 microm/3.5 degrees , 155 microm/6.8 degrees , 10.0%/7.0%, 0.91/0.89 in the light; and 64 microm/3.4 degrees , 124 microm/6.6 degrees , 9.9%/7.8% and 0.95/0.92 in the dark.
Conclusions: The anterior segment OCT demonstrated reliable anterior chamber angle measurement in different lighting conditions with good repeatability and reproducibility.
Conflict of interest statement
Competing interests: None declared.
Comment in
-
Anterior segment imaging for glaucoma: OCT or UBM?Br J Ophthalmol. 2007 Nov;91(11):1420-1. doi: 10.1136/bjo.2007.121038. Br J Ophthalmol. 2007. PMID: 17947263 Free PMC article.
References
-
- Pavlin C J, Harasiewicz K, Foster F S. Ultrasound biomicroscopy of anterior segment structures in normal and glaucomatous eyes. Am J Ophthalmol 1992113381–389. - PubMed
-
- Urbak S F, Pedersen J K, Thorsen T T. Ultrasound biomicroscopy. II. Intraobserver and interobserver reproducibility of measurements. Acta Ophthalmol Scand 199876546–549. - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources