. 2007 Nov;91(11):1490-2.
doi: 10.1136/bjo.2007.118901. Epub 2007 May 2.

Repeatability and reproducibility of anterior chamber angle measurement with anterior segment optical coherence tomography

Affiliations

Repeatability and reproducibility of anterior chamber angle measurement with anterior segment optical coherence tomography

Haitao Li et al. Br J Ophthalmol. 2007 Nov.

Abstract

Aim: To evaluate the repeatability and reproducibility of anterior chamber angle measurement obtained by anterior segment optical coherence tomography.

Methods: Twenty-five normal subjects were invited for anterior chamber angle imaging with an anterior segment optical coherence tomography (OCT) on one randomly selected eye in three separate visits within a week. Each eye was imaged three times under room light (light intensity = 368 lux) and three times in the dark during the first visit. In the subsequent visits, each eye was imaged once in the light and once in the dark. The angle opening distance (AOD 500) and the trabecular-iris angle (TIA 500) were measured by a single observer. Only the nasal angle was analysed. Intrasession and intersession within-subject standard deviation (Sw), precision (1.96xSw), coefficient of variation (CVw) (100xSw/overall mean), and intraclass correlation coefficient (ICC) were calculated to evaluate repeatability and reproducibility.

Results: For intrasession repeatability, the Sw, precision, CVw and ICC of AOD/TIA were 45 microm/2.4 degrees , 88 microm/4.7 degrees , 5.8%/4.8% and 0.97/0.95 in the light; and 45 microm/2.1 degrees , 88 microm/4.2 degrees , 7.0%/5.0% and 0.98/0.97 in the dark. For intersession reproducibility, the Sw, precision, CVw and ICC of AOD/TIA were 79 microm/3.5 degrees , 155 microm/6.8 degrees , 10.0%/7.0%, 0.91/0.89 in the light; and 64 microm/3.4 degrees , 124 microm/6.6 degrees , 9.9%/7.8% and 0.95/0.92 in the dark.

Conclusions: The anterior segment OCT demonstrated reliable anterior chamber angle measurement in different lighting conditions with good repeatability and reproducibility.

PubMed Disclaimer

Conflict of interest statement

Competing interests: None declared.

Comment in

References

    1. Pavlin C J, Harasiewicz K, Foster F S. Ultrasound biomicroscopy of anterior segment structures in normal and glaucomatous eyes. Am J Ophthalmol 1992113381–389. - PubMed
    1. Pavlin C J, Harasiewicz K, Sherar M D.et al Clinical use of ultrasound biomicroscopy. Ophthalmology 199198287–295. - PubMed
    1. Tello C, Liebmann J, Potash S D.et al Measurement of ultrasound biomicroscopy images: intraobserver and interobserver reliability. Invest Ophthalmol Vis Sci 1994353549–3552. - PubMed
    1. Urbak S F, Pedersen J K, Thorsen T T. Ultrasound biomicroscopy. II. Intraobserver and interobserver reproducibility of measurements. Acta Ophthalmol Scand 199876546–549. - PubMed
    1. Ishikawa H, Inazumi K, Liebmann J M.et al Inadvertent corneal indentation can cause artifactitious widening of the iridocorneal angle on ultrasound biomicroscopy. Ophthalmic Surg Lasers 200031342–345. - PubMed

Publication types