Differences in exhaled breath condensate pH measurements between samples obtained with two commercial devices
- PMID: 17481878
- DOI: 10.1016/j.rmed.2007.02.023
Differences in exhaled breath condensate pH measurements between samples obtained with two commercial devices
Abstract
Background: The objective of this study was to determine differences in EBC pH between samples obtained by RTube and EcoScreen, and to identify the effect of storage at -80 degrees C on the pH values.
Methods: Twenty-three nonsmoking subjects with asthma or allergic rhinitis or without respiratory disease performed two sequential exhaled breath condensate (EBC) collections, using the RTube collection system and the EcoScreen condenser. EBC pH was measured immediately after collection and after storage at -80 degrees C for 8 weeks, without deaeration and repeated following deaeration with Argon.
Results: In EBC samples without deaeration, the EcoScreen pH values were significantly higher than the RTube pH values (mean difference, 0.32; 95% CI, 0.21-0.44, P < 0.0001). In deaerated EBC samples, the EcoScreen pH values were also significantly higher than the RTube pH values (mean difference, 0.12; 95% CI, 0.01-0.25, P=0.04). For both EBC collection systems, storage for 8 weeks had a significant influence on pH of nondeaerated samples.
Conclusions: The present study demonstrates that EBC pH value is dependent on the collection device used and that the storage for 8 weeks had a significant influence on the pH of samples analyzed without deaeration.
Comment in
-
Effect of freezing on exhaled breath condensate pH.Respir Med. 2007 Dec;101(12):2566; author reply 2567. doi: 10.1016/j.rmed.2007.07.013. Epub 2007 Sep 24. Respir Med. 2007. PMID: 17890073 No abstract available.
Similar articles
-
Comparison of two methods for exhaled breath condensate collection.Allergy. 2006 Aug;61(8):1016-8. doi: 10.1111/j.1398-9995.2006.01064.x. Allergy. 2006. PMID: 16867057
-
Clinical and technical factors affecting pH and other biomarkers in exhaled breath condensate.Pediatr Pulmonol. 2006 Jan;41(1):87-94. doi: 10.1002/ppul.20296. Pediatr Pulmonol. 2006. PMID: 16292777
-
Ambient temperature impacts on pH of exhaled breath condensate.Respirology. 2010 Jan;15(1):155-9. doi: 10.1111/j.1440-1843.2009.01664.x. Epub 2009 Nov 23. Respirology. 2010. PMID: 19947991
-
Exhaled breath condensate in patients with asthma: implications for application in clinical practice.Clin Exp Allergy. 2008 Apr;38(4):557-65. doi: 10.1111/j.1365-2222.2008.02940.x. Clin Exp Allergy. 2008. PMID: 18352973 Review.
-
[Exhaled breath condensate in the assessment of airway inflammation].Przegl Lek. 2006;63(12):1321-5. Przegl Lek. 2006. PMID: 17642149 Review. Polish.
Cited by
-
Exhaled breath marker in asthma patients with gastroesophageal reflux disease.J Clin Biochem Nutr. 2007 Nov;41(3):147-53. doi: 10.3164/jcbn.2007020. J Clin Biochem Nutr. 2007. PMID: 18299708 Free PMC article.
-
Measures of asthma control.Curr Opin Pulm Med. 2012 Jan;18(1):48-56. doi: 10.1097/MCP.0b013e32834db0f2. Curr Opin Pulm Med. 2012. PMID: 22081089 Free PMC article. Review.
-
Lipoxin A(4) and 8-isoprostane in the exhaled breath condensate of children hospitalized for status asthmaticus.Pediatr Crit Care Med. 2012 Mar;13(2):141-5. doi: 10.1097/PCC.0b013e3182231644. Pediatr Crit Care Med. 2012. PMID: 21666533 Free PMC article.
-
Biomarkers in exhaled breath condensate: a review of collection, processing and analysis.J Breath Res. 2008 Sep;2(3):037004. doi: 10.1088/1752-7155/2/3/037004. Epub 2008 Sep 8. J Breath Res. 2008. PMID: 21386165 Free PMC article.
-
Comparison of exhaled breath condensate pH using two commercially available devices in healthy controls, asthma and COPD patients.Respir Res. 2009 Aug 24;10(1):78. doi: 10.1186/1465-9921-10-78. Respir Res. 2009. PMID: 19703285 Free PMC article. Clinical Trial.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical