Comparison of noninvasive oscillometric and intra-arterial blood pressure measurements in hyperacute stroke
- PMID: 17496464
- DOI: 10.1097/MBP.0b013e3280b083e2
Comparison of noninvasive oscillometric and intra-arterial blood pressure measurements in hyperacute stroke
Erratum in
- Blood Press Monit. 2007 Oct;12(5):349. Eleni, Koroboki [corrected to Koroboki, Eleni]
Abstract
Objectives: This study aims to compare automatic oscillometric blood pressure recordings with simultaneous direct intra-arterial blood pressure measurements in hyperacute stroke patients to test the accuracy of oscillometric readings.
Methods: A total of 51 first-ever stroke patients underwent simultaneous noninvasive automatic oscillometric and intra-arterial blood pressure monitoring within 3 h of ictus. Casual blood pressure was measured in both arms using a standard mercury sphygmomanometer on hospital admission. Patients who received antihypertensive medication during the blood pressure monitoring were excluded.
Results: The estimation of systolic blood pressure (SBP) using oscillometric recordings underestimated direct radial artery SBP by 9.7 mmHg (95% confidence interval: 6.5-13.0, P<0.001). In contrast, an upward bias of 5.6 mmHg (95% confidence interval: 3.5-7.7, P<0.001) was documented when noninvasive diastolic blood pressure (DBP) recordings were compared with intra-arterial DBP recordings. For SBP and DBP, the Pearson correlation coefficients between noninvasive and intra-arterial recordings were 0.854 and 0.832, respectively. When the study population was stratified according to SBP bands (group A: SBP<or=160 mmHg; group B: SBP>160 mmHg and SBP<or=180 mmHg, group C: SBP>180 mmHg), higher mean DeltaSBP (intra-arterial SBP-oscillometric SBP) levels were documented in group C (+19.8 mmHg, 95% confidence intervals: 12.2-27.4) when compared with groups B (+8.5 mmHg, 95% confidence intervals: 2.7-14.5; P=0.025) and A (+5.9 mmHg, 95% confidence intervals: 1.8-9.9; P=0.002).
Conclusion: Noninvasive automatic oscillometric BP measurements underestimate direct SBP recordings and overestimate direct DBP readings in acute stroke. The magnitude of the discrepancy between intra-arterial and oscillometric SBP recordings is even more prominent in patients with critically elevated SBP levels.
Similar articles
-
Random zero sphygmomanometer versus automatic oscillometric blood pressure monitor; is either the instrument of choice?J Hum Hypertens. 1995 Nov;9(11):885-9. J Hum Hypertens. 1995. PMID: 8583467 Clinical Trial.
-
A comparison of blood pressure and pulse pressure values obtained by oscillometric and central measurements in hypertensive patients.Blood Press. 2010 Apr;19(2):98-103. doi: 10.3109/08037050903516318. Blood Press. 2010. PMID: 20070247
-
Simultaneous validation of the SunTech 247 diagnostic station blood pressure measurement device according to the British Hypertension Society protocol, the International Protocol and the Association for the Advancement of Medical Instrumentation standards.Blood Press Monit. 2009 Oct;14(5):222-7. doi: 10.1097/MBP.0b013e328330c873. Blood Press Monit. 2009. PMID: 19741510
-
Blood pressure measurement: retrospective and prospective views.Am J Hypertens. 2011 Jun;24(6):628-34. doi: 10.1038/ajh.2011.22. Epub 2011 Feb 24. Am J Hypertens. 2011. PMID: 21350431 Review.
-
Can an automatic oscillometric device replace a mercury sphygmomanometer on blood pressure measurement? a systematic review and meta-analysis.Blood Press Monit. 2019 Dec;24(6):265-276. doi: 10.1097/MBP.0000000000000412. Blood Press Monit. 2019. PMID: 31658107
Cited by
-
Most patients with non-hypertensive diseases at a critical care resuscitation unit require arterial pressure monitoring: a prospective observational study.World J Emerg Med. 2023;14(3):173-178. doi: 10.5847/wjem.j.1920-8642.2023.035. World J Emerg Med. 2023. PMID: 37152525 Free PMC article.
-
Comparison of intraoperative blood pressure values measured by noninvasive versus invasive methods during normotension, hypertension, and hypotension.J Anaesthesiol Clin Pharmacol. 2024 Apr-Jun;40(2):258-263. doi: 10.4103/joacp.joacp_439_22. Epub 2023 Oct 18. J Anaesthesiol Clin Pharmacol. 2024. PMID: 38919432 Free PMC article.
-
A poor correlation exists between oscillometric and radial arterial blood pressure as measured by the Philips MP90 monitor.J Clin Monit Comput. 2009 Jun;23(3):169-74. doi: 10.1007/s10877-009-9178-8. Epub 2009 Apr 25. J Clin Monit Comput. 2009. PMID: 19396553
-
Increasing sleep duration to lower beat-to-beat blood pressure: a pilot study.J Sleep Res. 2013 Jun;22(3):295-304. doi: 10.1111/jsr.12011. Epub 2012 Nov 22. J Sleep Res. 2013. PMID: 23171375 Free PMC article. Clinical Trial.
-
Correlation of Noninvasive Blood Pressure and Invasive Intra-arterial Blood Pressure in Patients Treated with Vasoactive Medications in a Neurocritical Care Unit.Neurocrit Care. 2018 Jun;28(3):265-272. doi: 10.1007/s12028-018-0521-0. Neurocrit Care. 2018. PMID: 29569130
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical