Integrated siphon-based metering and sedimentation of whole blood on a hydrophilic lab-on-a-disk
- PMID: 17505885
- DOI: 10.1007/s10544-007-9076-0
Integrated siphon-based metering and sedimentation of whole blood on a hydrophilic lab-on-a-disk
Abstract
In this paper, we present a novel and fully integrated centrifugal microfluidic "lab-on-a-disk" for rapid colorimetric assays in human whole blood. All essential steps comprising blood sampling, metering, plasma extraction and the final optical detection are conducted within t=150 s in passive, globally hydrophilized structures which obviate the need for intricate local hydrophobic surface patterning. Our technology features a plasma extraction structure (V=500 nL, CV<5%) where the purified plasma (cRBC<0.11%) is centrifugally separated, metered by an overflow and subsequently extracted by a siphon-based principle through a hydrophilic extraction channel into the detection chamber.
Similar articles
-
Quantifying colorimetric assays in paper-based microfluidic devices by measuring the transmission of light through paper.Anal Chem. 2009 Oct 15;81(20):8447-52. doi: 10.1021/ac901307q. Anal Chem. 2009. PMID: 19722495
-
An integrated digital microfluidic lab-on-a-chip for clinical diagnostics on human physiological fluids.Lab Chip. 2004 Aug;4(4):310-5. doi: 10.1039/b403341h. Epub 2004 May 26. Lab Chip. 2004. PMID: 15269796
-
Sensitivity enhancement for colorimetric glucose assays on whole blood by on-chip beam-guidance.Biomed Microdevices. 2006 Sep;8(3):209-14. doi: 10.1007/s10544-006-8172-x. Biomed Microdevices. 2006. PMID: 16732473
-
Microfluidic platforms for lab-on-a-chip applications.Lab Chip. 2007 Sep;7(9):1094-110. doi: 10.1039/b706364b. Epub 2007 Jul 27. Lab Chip. 2007. PMID: 17713606 Review.
-
Fully integrated microfluidic separations systems for biochemical analysis.J Chromatogr A. 2007 Oct 19;1168(1-2):170-88; discussion 169. doi: 10.1016/j.chroma.2007.06.010. Epub 2007 Jun 12. J Chromatogr A. 2007. PMID: 17659293 Review.
Cited by
-
Systematic review of centrifugal valving based on digital twin modeling towards highly integrated lab-on-a-disc systems.Microsyst Nanoeng. 2021 Dec 16;7:104. doi: 10.1038/s41378-021-00317-3. eCollection 2021. Microsyst Nanoeng. 2021. PMID: 34987859 Free PMC article. Review.
-
Supernatant decanting on a centrifugal platform.Biomicrofluidics. 2011 Mar 30;5(1):13414. doi: 10.1063/1.3571477. Biomicrofluidics. 2011. PMID: 21522503 Free PMC article.
-
Rapid, culture-independent, optical diagnostics of centrifugally captured bacteria from urine samples.Biomicrofluidics. 2015 Aug 11;9(4):044118. doi: 10.1063/1.4928070. eCollection 2015 Jul. Biomicrofluidics. 2015. PMID: 26339318 Free PMC article.
-
Density-Gradient Mediated Band Extraction of Leukocytes from Whole Blood Using Centrifugo-Pneumatic Siphon Valving on Centrifugal Microfluidic Discs.PLoS One. 2016 May 11;11(5):e0155545. doi: 10.1371/journal.pone.0155545. eCollection 2016. PLoS One. 2016. PMID: 27167376 Free PMC article.
-
Microfluidics for drug discovery and development: from target selection to product lifecycle management.Drug Discov Today. 2008 Jan;13(1-2):1-13. doi: 10.1016/j.drudis.2007.10.003. Epub 2007 Nov 26. Drug Discov Today. 2008. PMID: 18190858 Free PMC article. Review.