Agreement among 3 methods to measure corneal thickness: ultrasound pachymetry, Orbscan II, and Visante anterior segment optical coherence tomography
- PMID: 17507097
- DOI: 10.1016/j.ophtha.2007.02.017
Agreement among 3 methods to measure corneal thickness: ultrasound pachymetry, Orbscan II, and Visante anterior segment optical coherence tomography
Abstract
Purpose: To assess the agreement of central and paracentral corneal thickness measurements between ultrasound pachymetry (USP), Orbscan II, and Visante anterior segment optical coherence tomography (ASOCT).
Design: Observational cross-sectional study.
Participants: Seventy eyes of 70 subjects.
Methods: Each subject underwent Orbscan II (using an acoustic equivalent correction factor of 0.89), ASOCT, and USP examination. Bland-Altman plots were used to evaluate agreement between instruments.
Main outcome measures: Central and paracentral corneal thickness measurements by the 3 methods and agreement, as evaluated by 95% limits of agreement (LOA).
Results: The mean measurements of average central corneal thickness by USP, Orbscan II, and ASOCT were 553.5+/-30.26 microm, 553.22+/-25.47 microm, and 538.79+/-26.22 microm, respectively. There was high correlation between instruments: USP with ASOCT (r = 0.936, P<0.001), USP with Orbscan II (r = 0.900, P<0.001) for central corneal thickness measurements, and Orbscan II with ASOCT for average paracentral 2- to 5-mm measurements (r = 0.947, P<0.001). The mean differences (and upper/lower LOA) for central corneal thickness measurements were 0.31+/-13.34 microm (26.44/-25.83) between USP and Orbscan II, 14.74+/-10.84 microm (36.0/-6.51) between USP and ASOCT, and 14.44+/-9.14 microm (32.36/-3.48) between Orbscan II and ASOCT. The average mean difference (and upper/lower LOA) between Orbscan II and ASOCT for paracentral 2- to 5-mm corneal thickness measurements was 10.35+/-8.67 microm (27.35/-6.65).
Conclusion: Anterior segment optical coherence tomography underestimated corneal thickness compared with that measured with USP. Anterior segment optical coherence tomography had better agreement with the gold standard USP, as compared with Orbscan II. However, important discrepancies among instruments exist. Clinicians should be aware that corneal thickness measurements are influenced by the method of measurement and that, although highly correlated, these instruments should not be used interchangeably for the assessment of corneal thickness.
Comment in
-
Corneal thickness.Ophthalmology. 2008 Dec;115(12):2315-6; author reply 2316-2316.e1. doi: 10.1016/j.ophtha.2008.08.021. Ophthalmology. 2008. PMID: 19041483 No abstract available.
Similar articles
-
Comparative study of central corneal thickness measurement with slit-lamp optical coherence tomography and visante optical coherence tomography.Ophthalmology. 2008 May;115(5):796-801.e2. doi: 10.1016/j.ophtha.2007.07.006. Epub 2007 Oct 4. Ophthalmology. 2008. PMID: 17916376
-
Effect of corneal thickness on the agreement between ultrasound and Orbscan II pachymetry.J Cataract Refract Surg. 2007 Oct;33(10):1694-700. doi: 10.1016/j.jcrs.2007.05.036. J Cataract Refract Surg. 2007. PMID: 17889762
-
High-speed optical coherence tomography of corneal opacities.Ophthalmology. 2007 Jul;114(7):1278-85. doi: 10.1016/j.ophtha.2006.10.033. Epub 2007 Feb 20. Ophthalmology. 2007. PMID: 17307254
-
Measuring the cornea: the latest developments in corneal topography.Curr Opin Ophthalmol. 2007 Jul;18(4):325-33. doi: 10.1097/ICU.0b013e3281ca7121. Curr Opin Ophthalmol. 2007. PMID: 17568210 Review.
-
[Precision and reliability of Orbscan and ultrasonic pachymetry].J Fr Ophtalmol. 2001 Nov;24(9):912-21. J Fr Ophtalmol. 2001. PMID: 11912833 Review. French.
Cited by
-
Central Corneal Thickness in Highly Myopic Eyes: Inter-device Agreement of Ultrasonic Pachymetry, Pentacam and Orbscan II Before and After Photorefractive Keratectomy.J Ophthalmic Vis Res. 2014 Jan;9(1):14-21. J Ophthalmic Vis Res. 2014. PMID: 24982727 Free PMC article.
-
Corneal topography from spectral optical coherence tomography (sOCT).Biomed Opt Express. 2011 Dec 1;2(12):3232-47. doi: 10.1364/BOE.2.003232. Epub 2011 Nov 4. Biomed Opt Express. 2011. PMID: 22162814 Free PMC article.
-
Comparison of central corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and the Artemis-2 VHF scanner in normal eyes.Clin Ophthalmol. 2012;6:1037-43. doi: 10.2147/OPTH.S32955. Epub 2012 Jul 6. Clin Ophthalmol. 2012. PMID: 22848145 Free PMC article.
-
Comparison of central corneal thickness: ultrasound pachymetry versus slit-lamp optical coherence tomography, specular microscopy, and Orbscan.Clin Ophthalmol. 2015 Jun 12;9:1065-70. doi: 10.2147/OPTH.S81376. eCollection 2015. Clin Ophthalmol. 2015. PMID: 26109840 Free PMC article.
-
[Difference and distance between the central and thinnest points of the cornea: impact of refractive state, age and ocular side].Ophthalmologe. 2014 Apr;111(4):339-47. doi: 10.1007/s00347-013-2892-0. Ophthalmologe. 2014. PMID: 23921813 German.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources