Screening mammography-detected cancers: sensitivity of a computer-aided detection system applied to full-field digital mammograms
- PMID: 17507722
- DOI: 10.1148/radiol.2441060756
Screening mammography-detected cancers: sensitivity of a computer-aided detection system applied to full-field digital mammograms
Abstract
Purpose: To retrospectively evaluate the sensitivity of the performance of a computer-aided detection (CAD) system applied to full-field digital mammograms for detection of breast cancers in a screening group, with histologic findings as the reference standard.
Materials and methods: This study had institutional review board approval, and patient informed consent was waived. A commercially available CAD system was applied to the digital mammograms of 103 women (mean age, 51 years; range, 35-69 years) with 103 breast cancers detected with screening. Sensitivity values of the CAD system according to mammographic appearance, breast composition, and histologic findings were analyzed. Normal mammograms from 100 women (mean age, 54 years; age range, 35-75 years) with no mammographic and clinical abnormality during 2-year follow-up were used to determine false-positive CAD system marks. Differences between the cancer detection rates in fatty and dense breasts for the CAD system were compared by using the chi(2) test.
Results: The CAD system correctly marked 99 (96.1%) of 103 breast cancers. The CAD system marked all 44 breast cancers that manifested as microcalcifications only, all 23 breast cancers that manifested as a mass with microcalcifications, and 32 (89%) of 36 lesions that appeared as a mass only. The sensitivity of the CAD system in the fatty breast group was 95% (59 of 62) and in the dense breast group was 98% (40 of 41) (P = .537). The CAD system correctly marked all 31 lesions of ductal carcinoma in situ (DCIS), all 22 lesions of invasive ductal carcinoma with DCIS, the single invasive lobular carcinoma lesion, and 45 (92%) of 49 lesions of invasive ductal carcinoma. On normal mammograms, the mean number of false-positive marks per patient was 1.80 (range, 0-10 marks; median, 1 mark).
Conclusion: The CAD system can correctly mark most (96.1%) asymptomatic breast cancers detected with digital mammographic screening, with acceptable false-positive marks (1.80 per patient).
(c) RSNA, 2007.
Similar articles
-
Computer-aided detection in full-field digital mammography: sensitivity and reproducibility in serial examinations.Radiology. 2008 Jan;246(1):71-80. doi: 10.1148/radiol.2461062072. Radiology. 2008. PMID: 18096530
-
[Impact of breast density on computer-aided detection (CAD) of breast cancer].Zhonghua Zhong Liu Za Zhi. 2012 May;34(5):360-3. doi: 10. 3760/cma.j.issn.0253-3766.2012.05.009. Zhonghua Zhong Liu Za Zhi. 2012. PMID: 22883457 Chinese.
-
Evaluation of breast cancer with a computer-aided detection system by mammographic appearance and histopathology.Cancer. 2005 Sep 1;104(5):931-5. doi: 10.1002/cncr.21255. Cancer. 2005. PMID: 16047331
-
[Understanding CAD (computer-aided diagnosis) in mammography].J Radiol. 2005 Jan;86(1):29-35. doi: 10.1016/s0221-0363(05)81319-8. J Radiol. 2005. PMID: 15785414 Review. French.
-
Recent advances in breast-specific imaging.Cancer Control. 2007 Oct;14(4):338-49. doi: 10.1177/107327480701400404. Cancer Control. 2007. PMID: 17914334 Review.
Cited by
-
Computer-aided detection of breast masses depicted on full-field digital mammograms: a performance assessment.Br J Radiol. 2012 Jun;85(1014):e153-61. doi: 10.1259/bjr/51461617. Epub 2011 Feb 22. Br J Radiol. 2012. PMID: 21343322 Free PMC article.
-
Detection of breast cancer with a computer-aided detection applied to full-field digital mammography.J Digit Imaging. 2013 Aug;26(4):768-73. doi: 10.1007/s10278-012-9564-5. J Digit Imaging. 2013. PMID: 23319110 Free PMC article.
-
Automated feature set selection and its application to MCC identification in digital mammograms for breast cancer detection.Sensors (Basel). 2013 Apr 11;13(4):4855-75. doi: 10.3390/s130404855. Sensors (Basel). 2013. PMID: 23580053 Free PMC article.
-
Role of computer-aided detection in very small screening detected invasive breast cancers.J Digit Imaging. 2013 Jun;26(3):572-7. doi: 10.1007/s10278-012-9550-y. J Digit Imaging. 2013. PMID: 23131867 Free PMC article.
-
Computer-aided detection of breast carcinoma in standard mammographic projections with digital mammography.Int J Comput Assist Radiol Surg. 2009 Jun;4(4):331-6. doi: 10.1007/s11548-009-0300-7. Epub 2009 Apr 15. Int J Comput Assist Radiol Surg. 2009. PMID: 20033580