High-resolution whole-body magnetic resonance imaging applications at 1.5 and 3 Tesla: a comparative study
- PMID: 17507818
- DOI: 10.1097/01.rli.0000262089.55618.18
High-resolution whole-body magnetic resonance imaging applications at 1.5 and 3 Tesla: a comparative study
Abstract
Objectives: To analyze the impact of altered magnetic field properties on image quality and on potential artifacts when an established whole-body magnetic resonance imaging (WB-MRI) protocol at 1.5 Tesla (T) is migrated to 3 T.
Materials and methods: Fifteen volunteers underwent noncontrast magnetic resonance imaging (MRI) on 32-channel whole body-scanners at 1.5 and 3 T with the use of parallel acquisition techniques (PAT). Coronal T1-weighted TSE- and short tau inversion recovery (STIR)-sequences at 4 body levels including sagittal imaging of the whole spine were performed. Additional axial HASTE-imaging of lung and abdomen, T1-/T2-weighted-TSE- and EPI-sequences of the brain and T2-weighted respiratory-triggered imaging of the liver was acquired. Both data sets were compared by 2 independent readers in respect to artifacts and image quality using a 5-point scale. Regions of pronounced artifacts were defined.
Results: Overall image impression was both qualitatively rated as "good" at 1.5 and 3 T for T1-w-TSE- and STIR-imaging of the whole body and spine. At 1.5 T, significantly better quantitative values for overall image quality were found for WB-STIR, T2-w-TSE imaging of the liver and brain (Wilcoxon Mann-Whitney U Test; P < 0.05), overall rated as good at 3 T. Significantly higher dielectric effects at 3 T were affecting T1-w- and STIR-WB-MRI, and HASTE of the abdomen and better image homogeneity at 1.5 T was observed for T1-weighted-/STIR-WB-MRI and T1-w-TSE-imaging of the spine. Pulsation artifacts were significantly increased at 3 T for T1-w WB-MRI. Significantly higher susceptibility artifacts were found for GRE-sequences of the brain at 3 T. Motion artifacts, Gibbs-Ringing, and image distortion was not significantly different and showed slightly higher quantitative values at 3 T (except for HASTE imaging of the abdomen). Overall scan time was 45 minutes and 44 seconds at 1.5 T and 40 minutes and 28 seconds at 3 T at identical image resolution.
Conclusion: Three Tesla WB-MRI is feasible with good image quality comparable to 1.5 T. 3.0 T WB-MRI shows significantly more artifacts with a mild to moderate impact on image assessment. Therefore 1.5 T WB-MRI is the preferred image modality. Overall scan time at 3 T is reduced with the use of parallel imaging at a constant image resolution.
Similar articles
-
Magnetic resonance imaging of the upper abdomen using a free-breathing T2-weighted turbo spin echo sequence with navigator triggered prospective acquisition correction.J Magn Reson Imaging. 2005 May;21(5):576-82. doi: 10.1002/jmri.20293. J Magn Reson Imaging. 2005. PMID: 15834908
-
Brain magnetic resonance imaging at 3 Tesla using BLADE compared with standard rectilinear data sampling.Invest Radiol. 2006 Jul;41(7):586-92. doi: 10.1097/01.rli.0000223742.35655.24. Invest Radiol. 2006. PMID: 16772852
-
Lung MRI at 1.5 and 3 Tesla: observer preference study and lesion contrast using five different pulse sequences.Invest Radiol. 2007 Jun;42(6):377-83. doi: 10.1097/01.rli.0000261926.86278.96. Invest Radiol. 2007. PMID: 17507808
-
Whole-body magnetic resonance imaging and positron emission tomography-computed tomography in oncology.Top Magn Reson Imaging. 2007 Jun;18(3):193-202. doi: 10.1097/RMR.0b013e318093e6bo. Top Magn Reson Imaging. 2007. PMID: 17762383 Review.
-
Whole-body MRI for the staging and follow-up of patients with metastasis.Eur J Radiol. 2009 Jun;70(3):393-400. doi: 10.1016/j.ejrad.2009.03.045. Epub 2009 May 19. Eur J Radiol. 2009. PMID: 19457631 Review.
Cited by
-
Comparison of spinal anatomy between 3-Tesla MRI and CT-myelography under healthy and pathological conditions.Surg Radiol Anat. 2010 Jul;32(6):581-5. doi: 10.1007/s00276-009-0601-0. Epub 2009 Dec 6. Surg Radiol Anat. 2010. PMID: 19967536
-
Renal imaging at 5 T versus 3 T: a comparison study.Insights Imaging. 2022 Sep 24;13(1):155. doi: 10.1186/s13244-022-01290-9. Insights Imaging. 2022. PMID: 36153471 Free PMC article.
-
Comprehensive imaging of residual/ recurrent nasopharyngeal carcinoma using whole-body MRI at 3 T compared with FDG-PET-CT.Eur Radiol. 2010 Sep;20(9):2229-40. doi: 10.1007/s00330-010-1784-9. Eur Radiol. 2010. PMID: 20393714
-
1.5 versus 3 versus 7 Tesla in abdominal MRI: A comparative study.PLoS One. 2017 Nov 10;12(11):e0187528. doi: 10.1371/journal.pone.0187528. eCollection 2017. PLoS One. 2017. PMID: 29125850 Free PMC article.
-
Value of IVIM in Differential Diagnoses between Benign and Malignant Solitary Lung Nodules and Masses: A Meta-analysis.Front Surg. 2022 Jun 1;9:817443. doi: 10.3389/fsurg.2022.817443. eCollection 2022. Front Surg. 2022. PMID: 36017515 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical