Device monitoring of intrathoracic impedance: clinical observations from a patient registry
- PMID: 17512419
- DOI: 10.1016/j.amjcard.2007.02.039
Device monitoring of intrathoracic impedance: clinical observations from a patient registry
Abstract
A distinct advantage of implantable device diagnostics is that the data may be made available on a continuous basis, rather than at sporadic intervals associated with clinical testing. Recently, intrathoracic impedance monitoring has also become available in some implantable devices as an index of congestion and thoracic fluid accumulation secondary to decompensated heart failure. Despite the potential advantages, new advances in implantable device diagnostic technology also pose important questions regarding their clinical application. To examine these questions and to generate hypotheses, the relation between daily intrathoracic impedance measurements and other physical measurements or comorbidities was studied. A retrospective review was performed of 25 patients who previously underwent implantation of a cardiac resynchronization therapy/defibrillator device with the capability to continuously monitor intrathoracic impedance. This limited scope analysis demonstrated that daily measurement of intrathoracic impedances might reveal the intrinsic relations between heart failure decompensation and the onset of atrial and ventricular arrhythmias. Abnormal patterns of intrathoracic impedance that has increased and plateaued after implant may indicate worsening heart failure. The severity of congestive heart failure at the time of interrogation may correlate with device-based impedance measurements. Weight, sex, and body index may have a limited impact on impedance, and the lowest impedances may be seen in older patients. Patients with severe pulmonary disease may present with unique daily impedance profiles. Finally, daily impedance may have unpredictable relations with other clinical markers of heart failure. In summary, intrathoracic impedance represents a clinically useful diagnostic tool that can increase our understanding of a dynamic disease state on an individual patient basis.
Similar articles
-
Changes in intrathoracic impedance are associated with subsequent risk of hospitalizations for acute decompensated heart failure: clinical utility of implanted device monitoring without a patient alert.J Card Fail. 2009 Aug;15(6):475-81. doi: 10.1016/j.cardfail.2009.01.012. Epub 2009 Mar 17. J Card Fail. 2009. PMID: 19643357
-
Integrating device-based monitoring into clinical practice: insights from a large heart failure clinic.Am J Cardiol. 2007 May 21;99(10A):17G-22G. doi: 10.1016/j.amjcard.2007.02.038. Epub 2007 Mar 9. Am J Cardiol. 2007. PMID: 17512418 Review.
-
Fundamentals of intrathoracic impedance monitoring in heart failure.Am J Cardiol. 2007 May 21;99(10A):3G-10G. doi: 10.1016/j.amjcard.2007.02.009. Epub 2007 Mar 9. Am J Cardiol. 2007. PMID: 17512421 Review.
-
Use of device diagnostics as an educational tool to improve patient adherence.Am J Cardiol. 2007 May 21;99(10A):29G-33G. doi: 10.1016/j.amjcard.2007.02.042. Epub 2007 Mar 12. Am J Cardiol. 2007. PMID: 17512420 Review.
-
Continuous monitoring of intrathoracic impedance and right ventricular pressures in patients with heart failure.Circ Heart Fail. 2010 May;3(3):370-7. doi: 10.1161/CIRCHEARTFAILURE.109.867549. Epub 2010 Mar 2. Circ Heart Fail. 2010. PMID: 20197559 Clinical Trial.
Cited by
-
Threshold crossing of device-based intrathoracic impedance trends identifies relatively increased mortality risk.Eur Heart J. 2012 Sep;33(17):2189-96. doi: 10.1093/eurheartj/ehs121. Epub 2012 Jun 7. Eur Heart J. 2012. PMID: 22677137 Free PMC article.
-
Influence of lung volume, fluid and capillary recruitment during positional changes and exercise on thoracic impedance in heart failure.Respir Physiol Neurobiol. 2014 Oct 1;202:75-81. doi: 10.1016/j.resp.2014.08.001. Epub 2014 Aug 14. Respir Physiol Neurobiol. 2014. PMID: 25128641 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical