Efficacy of Cruise Control in controlling postocclusion surge with Legacy and Millennium venturi phacoemulsification machines
- PMID: 17531704
- DOI: 10.1016/j.jcrs.2007.02.028
Efficacy of Cruise Control in controlling postocclusion surge with Legacy and Millennium venturi phacoemulsification machines
Abstract
Purpose: To determine the efficacy of the Cruise Control surge-limiting device (Staar Surgical) with phacoemulsification machines known to have high levels of surge.
Setting: John A. Moran Eye Center Clinical Laboratories.
Methods: In an in vitro study, postocclusion anterior chamber depth changes were measured in fresh phakic human eye-bank eyes using the Alcon Legacy and Bausch & Lomb Millennium venturi machines in conjunction with the Staar Cruise Control device. Both machines were tested with 19-gauge non-Aspiration Bypass System tips at high-surge settings (500 mm Hg vacuum pressure, 75 cm bottle height, 40 mL/min flow rate for the Legacy) and low-surge settings (400 mm Hg vacuum pressure, 125 cm bottle height, 40 mL/min flow rate for the Legacy). Adjusted parameters of flow, vacuum, and irrigation were used based on previous studies to create identical conditions for each device tested. The effect of the Cruise Control device on aspiration rates was also tested with both machines at the low-surge settings.
Results: At the high setting with the addition of Cruise Control, surge decreased significantly with the Legacy but was too large to measure with the Millennium venturi. At the low setting with the addition of Cruise Control, surge decreased significantly with both machines. Surge with the Millennium decreased from more than 1.0 mm to a mean of 0.21 mm +/- 0.02 (SD) (P<.0001). Surge with the Legacy decreased from a mean of 0.09 +/- 0.02 mm to 0.05 +/- 0 mm, a 42.9% decrease (P<.0001). The Millennium had the highest surge and aspiration rate before Cruise Control and the greatest percentage decrease in the surge and aspiration rates as a result of the addition of Cruise Control.
Conclusions: In the Legacy machine, the Cruise Control device had a statistically and clinically significant effect. Cruise Control had a large effect on fluidics as well as surge amplitude with the Millennium machine. The greater the flow or greater the initial surge, the greater the impact of the Cruise Control device.
Similar articles
-
Objective measurement of postocclusion surge during phacoemulsification in human eye-bank eyes.Am J Ophthalmol. 2007 Mar;143(3):437-40. doi: 10.1016/j.ajo.2006.11.017. Epub 2006 Dec 18. Am J Ophthalmol. 2007. PMID: 17222793
-
Effect of bottle height and aspiration rate on postocclusion surge in Infiniti and Millennium peristaltic phacoemulsification machines.J Cataract Refract Surg. 2008 Aug;34(8):1400-2. doi: 10.1016/j.jcrs.2008.04.042. J Cataract Refract Surg. 2008. PMID: 18655995
-
Fluidics and heat generation of Alcon Infiniti and Legacy, Bausch & Lomb Millennium, and advanced medical optics sovereign phacoemulsification systems.Am J Ophthalmol. 2006 Sep;142(3):387-92. doi: 10.1016/j.ajo.2006.04.042. Am J Ophthalmol. 2006. PMID: 16935581
-
Phaco fluidics and phaco ultrasound power modulations.Ophthalmol Clin North Am. 2006 Dec;19(4):457-68. doi: 10.1016/j.ohc.2006.08.001. Ophthalmol Clin North Am. 2006. PMID: 17067901 Review.
-
Phacoemulsification in review: Optimization of cataract removal in an in vitro setting.Surv Ophthalmol. 2019 Nov-Dec;64(6):868-875. doi: 10.1016/j.survophthal.2019.06.007. Epub 2019 Jul 2. Surv Ophthalmol. 2019. PMID: 31276738 Review.
MeSH terms
LinkOut - more resources
Full Text Sources