Reliability of temperatures measured at standard monitoring sites as an index of brain temperature during deep hypothermic cardiopulmonary bypass conducted for thoracic aortic reconstruction
- PMID: 17532957
- DOI: 10.1016/j.jtcvs.2006.11.031
Reliability of temperatures measured at standard monitoring sites as an index of brain temperature during deep hypothermic cardiopulmonary bypass conducted for thoracic aortic reconstruction
Abstract
Objective: It is essential to estimate the brain temperature of patients during deliberate deep hypothermia. Using jugular bulb temperature as a standard for brain temperature, we evaluated the accuracy and precision of 5 standard temperature monitoring sites (ie, pulmonary artery, nasopharynx, forehead deep-tissue, urinary bladder, and fingertip skin-surface tissue) during deep hypothermic cardiopulmonary bypass conducted for thoracic aortic reconstruction.
Methods: In 20 adult patients with thoracic aortic aneurysms, the 5 temperature monitoring sites were recorded every 1 minute during deep hypothermic (<20 degrees C) cardiopulmonary bypass. The accuracy was evaluated by the difference from jugular bulb temperature, and the precision was evaluated by its standard deviation, as well as by the correlation with jugular bulb temperature.
Results: Pulmonary artery temperature and jugular bulb temperature began to change immediately after the start of cooling or rewarming, closely matching each other, and the other temperatures lagged behind these two temperatures. During either situation, the accuracy of pulmonary artery temperature measurement (0.3 degrees C-0.5 degrees C) was much superior to the other measurements, and its precision (standard deviation of the difference from jugular bulb temperature = 1.5 degrees C-1.8 degrees C; correlation coefficient = 0.94-0.95) was also best among the measurements, with its rank order being pulmonary artery > or = nasopharynx > forehead > bladder > fingertip. However, the accuracy and precision of pulmonary artery temperature measurement was significantly impaired during and for several minutes after infusion of cold cardioplegic solution.
Conclusions: Pulmonary artery temperature measurement is recommended to estimate brain temperature during deep hypothermic cardiopulmonary bypass, even if it is conducted with the sternum opened; however, caution needs to be exercised in interpreting its measurements during periods of the cardioplegic solution infusion.
Similar articles
-
Changes in body temperature during profound hypothermic cardiopulmonary bypass in adult patients undergoing aortic arch reconstruction.J Anesth. 2004;18(2):73-81. doi: 10.1007/s00540-003-0225-1. J Anesth. 2004. PMID: 15127253
-
Temperature monitoring during cardiopulmonary bypass--do we undercool or overheat the brain?Eur J Cardiothorac Surg. 2004 Sep;26(3):580-5. doi: 10.1016/j.ejcts.2004.05.004. Eur J Cardiothorac Surg. 2004. PMID: 15302054
-
Accuracy of core temperature measurement in deep hypothermic circulatory arrest.Interact Cardiovasc Thorac Surg. 2008 Oct;7(5):922-4. doi: 10.1510/icvts.2008.181974. Epub 2008 Jul 25. Interact Cardiovasc Thorac Surg. 2008. PMID: 18658167
-
The importance of acid-base management for cardiac and cerebral preservation during open heart operations.Surg Gynecol Obstet. 1984 Apr;158(4):391-414. Surg Gynecol Obstet. 1984. PMID: 6424251 Review.
-
Cardiopulmonary bypass temperature and brain function.Anaesthesia. 2005 Apr;60(4):365-72. doi: 10.1111/j.1365-2044.2005.04112.x. Anaesthesia. 2005. PMID: 15766340 Review.
Cited by
-
Is Esophageal Temperature Better to Estimate Brain Temperature during Target Temperature Management in a Porcine Model of Cardiopulmonary Resuscitation?Biomed Res Int. 2017;2017:1279307. doi: 10.1155/2017/1279307. Epub 2017 Dec 20. Biomed Res Int. 2017. PMID: 29423402 Free PMC article.
-
Accuracy and precision of four common peripheral temperature measurement methods in intensive care patients.Med Devices (Auckl). 2016 Sep 1;9:301-8. doi: 10.2147/MDER.S109904. eCollection 2016. Med Devices (Auckl). 2016. PMID: 27621673 Free PMC article.
-
Targeted temperature management: Current evidence and practices in critical care.Indian J Crit Care Med. 2015 Sep;19(9):537-46. doi: 10.4103/0972-5229.164806. Indian J Crit Care Med. 2015. PMID: 26430341 Free PMC article. Review.
-
Current conduct of deep hypothermic circulatory arrest in China.HSR Proc Intensive Care Cardiovasc Anesth. 2013;5(1):25-32. HSR Proc Intensive Care Cardiovasc Anesth. 2013. PMID: 23734286 Free PMC article.
-
Insight into the use of tympanic temperature during target temperature management in emergency and critical care: a scoping review.J Intensive Care. 2021 Jun 12;9(1):43. doi: 10.1186/s40560-021-00558-4. J Intensive Care. 2021. PMID: 34118993 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical