. 2007 Aug;18(8):1375-86.
doi: 10.1016/j.jasms.2007.04.019. Epub 2007 Apr 29.

Pulsed large volume injection gas chromatography coupled with electron-capture negative ionization quadrupole mass spectrometry for simultaneous determination of typical halogenated persistent organic pollutants

Affiliations
Free article

Pulsed large volume injection gas chromatography coupled with electron-capture negative ionization quadrupole mass spectrometry for simultaneous determination of typical halogenated persistent organic pollutants

Yuli Zhao et al. J Am Soc Mass Spectrom. 2007 Aug.
Free article

Abstract

A pulsed large-volume injection gas chromatography coupled with electron-capture negative ionization quadrupole mass spectrometry (pLVI-GC/ECNI-qMS) was developed for the simultaneous determination of typical halogenated persistent organic pollutants (H-POPs). By monitoring the characteristic ions of large mass-to-charge ratio (m/z) for each of the H-POPs rather than the chlorine and/or bromine ions, this method avoided the possible interferences arising from the H-POPs themselves and from complex matrices encountered frequently in current GC/qMS methods; and allowed, on the other hand, the use of (13)C-labeled and perdeuterated analogues as internal standards for reliable quantification. pLVI up to 120 microL improved the instrumental detection limits down to pg-fg mL(-1), comparable to or lower than those obtained by the recognized GC/high-resolution MS methods reported so far. The H-POPs including 12 polybrominated diphenyl ethers, 1 polybrominated biphenyl, 10 polychlorinated biphenyls (PCBs), 4 hexachlorocyclohexane isomers, and hexachlorobenzene were involved in this study. The method developed demonstrated good linearity (r(2) = 0.9904-0.9999) within 0.5 to 50,000 pg mL(-1) for PCBs and 0.05 to 5000 pg mL(-1) for other H-POPs, and was satisfactory in terms of both repeatability (0.07%-2.2%) and reproducibility (2.1%-8.4%). It was validated by analyzing a NIST standard reference material SRM-1946 of Lake Superior fish tissue with low 0.01 to 63 pg g(-1) method detection limits, and successfully applied to the determination of the H-POPs in five reference materials of different matrices.

PubMed Disclaimer

References

    1. J Chromatogr A. 2003 Jun 6;1000(1-2):125-51 - PubMed
    1. J Mass Spectrom. 2003 Apr;38(4):394-400 - PubMed
    1. Anal Chem. 2003 Dec 1;75(23):6696-700 - PubMed
    1. Anal Chem. 2005 Dec 1;77(23):7662-72 - PubMed
    1. Anal Chem. 2002 Sep 1;74(17):4513-8 - PubMed

Publication types

MeSH terms

Substances

LinkOut - more resources