Degenerate mode band-pass birdcage coil for accelerated parallel excitation
- PMID: 17534905
- DOI: 10.1002/mrm.21247
Degenerate mode band-pass birdcage coil for accelerated parallel excitation
Abstract
An eight-rung, 3T degenerate birdcage coil (DBC) was constructed and evaluated for accelerated parallel excitation of the head with eight independent excitation channels. Two mode configurations were tested. In the first, each of the eight loops formed by the birdcage was individually excited, producing an excitation pattern similar to a loop coil array. In the second configuration a Butler matrix transformed this "loop coil" basis set into a basis set representing the orthogonal modes of the birdcage coil. In this case the rung currents vary sinusoidally around the coil and only four of the eight modes have significant excitation capability (the other four produce anticircularly polarized (ACP) fields). The lowest useful mode produces the familiar uniform B(1) field pattern, and the higher-order modes produce center magnitude nulls and azimuthal phase variations. The measured magnitude and phase excitation profiles of the individual modes were used to generate one-, four-, six-, and eightfold-accelerated spatially tailored RF excitations with 2D and 3D k-space excitation trajectories. Transmit accelerations of up to six-fold were possible with acceptable levels of spatial artifact. The orthogonal basis set provided by the Butler matrix was found to be advantageous when an orthogonal subset of these modes was used to mitigate B(1) transmit inhomogeneities using parallel excitation.
Similar articles
-
Parallel RF transmission with eight channels at 3 Tesla.Magn Reson Med. 2006 Nov;56(5):1163-71. doi: 10.1002/mrm.21042. Magn Reson Med. 2006. PMID: 17036289
-
Slice-selective RF pulses for in vivo B1+ inhomogeneity mitigation at 7 tesla using parallel RF excitation with a 16-element coil.Magn Reson Med. 2008 Dec;60(6):1422-32. doi: 10.1002/mrm.21739. Magn Reson Med. 2008. PMID: 19025908 Free PMC article.
-
Multiphoton parallel transmission (MP-pTx): Pulse design methods and numerical validation.Magn Reson Med. 2024 Oct;92(4):1376-1391. doi: 10.1002/mrm.30116. Epub 2024 Jun 20. Magn Reson Med. 2024. PMID: 38899391
-
Novel quadrature birdcage coil for a vertical B(0) field open MRI system.Magn Reson Med. 2000 Oct;44(4):633-40. doi: 10.1002/1522-2594(200010)44:4<633::aid-mrm18>3.0.co;2-k. Magn Reson Med. 2000. PMID: 11025520
-
Generalized encoding through the use of selective excitation in accelerated parallel MRI.NMR Biomed. 2006 May;19(3):379-92. doi: 10.1002/nbm.1047. NMR Biomed. 2006. PMID: 16705639 Review.
Cited by
-
Accelerating the co-simulation method for the design of transmit array coils for MRI.MAGMA. 2021 Feb;34(1):165-178. doi: 10.1007/s10334-020-00858-0. Epub 2020 Jun 27. MAGMA. 2021. PMID: 32594275
-
Recent Progress in Birdcage RF Coil Technology for MRI System.Diagnostics (Basel). 2020 Nov 27;10(12):1017. doi: 10.3390/diagnostics10121017. Diagnostics (Basel). 2020. PMID: 33261167 Free PMC article. Review.
-
Tailored excitation using nonlinear B0-shims.Magn Reson Med. 2012 Mar;67(3):601-8. doi: 10.1002/mrm.23278. Epub 2011 Dec 29. Magn Reson Med. 2012. PMID: 22222623 Free PMC article.
-
Four-dimensional spectral-spatial RF pulses for simultaneous correction of B1+ inhomogeneity and susceptibility artifacts in T2*-weighted MRI.Magn Reson Med. 2010 Jul;64(1):1-8. doi: 10.1002/mrm.22471. Magn Reson Med. 2010. PMID: 20577982 Free PMC article.
-
Specific absorption rate studies of the parallel transmission of inner-volume excitations at 7T.J Magn Reson Imaging. 2008 Oct;28(4):1005-18. doi: 10.1002/jmri.21548. J Magn Reson Imaging. 2008. PMID: 18821601 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical