Detection of proximal caries with high-resolution and standard resolution digital radiographic systems
- PMID: 17536087
- DOI: 10.1259/dmfr/99904877
Detection of proximal caries with high-resolution and standard resolution digital radiographic systems
Erratum in
- Dentomaxillofac Radiol. 2007 Sep;36(6):374-5
Abstract
Aims: The aim of this study was to: (1) compare the diagnostic accuracy of the high-resolution and standard resolution settings of four digital imaging systems for caries diagnosis and (2) compare the effect on the diagnostic accuracy of reducing the high-resolution image sizes to the standard resolution dimensions, and vice versa.
Materials and methods: 90 extracted human premolars were mounted in groups of 5 in plaster blocks, containing 4 test teeth and 1 non-test tooth. Two blocks at a time were placed in a jig to simulate a bitewing radiograph. Radiographs were taken using four digital systems (Planmeca Dixi 2; Gendex Visualix HDI; Dürr Vistascan; Digora Optime), each at two resolution settings. Next, the teeth were sectioned and a total of 65 surfaces were incorporated in the study. Additionally, the bicubic interpolation method was applied to reduce the high-resolution original images and to enlarge the standard resolution images. The original, reduced and enlarged images were randomly shown to five observers in two random sessions. The observers were asked to assess caries depth on a 4-point scale. The observers' scores were compared with the results from a histological examination. Data were analysed using the statistical theory for multivariate discrete data. Cohen's kappa was used to determine the agreement with the gold standard.
Results: None of the comparisons between the spatial resolution settings, or the comparisons between increased or reduced image size and the original image sizes, showed significant differences in the probability of caries detection (chi2=26.59, df=26, P approximately 0.50). The four digital systems used in this study differ significantly in the probability of caries detection (chi2=41.55, df=24, P<0.02). Compared with the gold standard, the Gendex Visualix HDI CCD sensor has the highest measure of agreement (kappa=0.31).
Conclusion: Caries diagnosis does not improve when using high-resolution settings compared with the standard settings. The use of bicubic convolution interpolation for zooming has no detectable effect on caries diagnosis and therefore is recommended to use when enlarging or reducing radiographs. The probability of caries detection, however, was different for the sensor systems of the different manufacturers.
Similar articles
-
Detection of proximal caries using digital radiographic systems with different resolutions.Indian J Dent Res. 2015 Jan-Feb;26(1):5-10. doi: 10.4103/0970-9290.156787. Indian J Dent Res. 2015. PMID: 25961607
-
Influence of spatial resolution and bit depth on detection of small caries lesions with digital receptors.Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2007 Mar;103(3):418-22. doi: 10.1016/j.tripleo.2006.05.016. Epub 2006 Aug 2. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2007. PMID: 17321456
-
Diagnostic accuracy of two software modalities for detection of caries lesions in digital radiographs from four dental systems.Dentomaxillofac Radiol. 2006 Mar;35(2):78-82. doi: 10.1259/dmfr/50356588. Dentomaxillofac Radiol. 2006. PMID: 16549433
-
Digital radiography and caries diagnosis.Dentomaxillofac Radiol. 1998 Jan;27(1):3-11. doi: 10.1038/sj.dmfr.4600321. Dentomaxillofac Radiol. 1998. PMID: 9482015 Review.
-
A review of dentists' use of digital radiography and caries diagnosis with digital systems.Dentomaxillofac Radiol. 2006 Sep;35(5):307-14. doi: 10.1259/dmfr/64693712. Dentomaxillofac Radiol. 2006. PMID: 16940477 Review.
Cited by
-
The interference of digital radiographic image acquisition and processing protocols in the diagnosis of incipient enamel carious lesions.J Clin Exp Dent. 2024 Feb 1;16(2):e96-e103. doi: 10.4317/jced.60990. eCollection 2024 Feb. J Clin Exp Dent. 2024. PMID: 38496805 Free PMC article.
-
Imaging modalities to inform the detection and diagnosis of early caries.Cochrane Database Syst Rev. 2021 Mar 15;3(3):CD014545. doi: 10.1002/14651858.CD014545. Cochrane Database Syst Rev. 2021. PMID: 33720395 Free PMC article.
-
Intraoral versus extraoral bitewing radiography in detection of enamel proximal caries: an ex vivo study.Dentomaxillofac Radiol. 2016;45(4):20150326. doi: 10.1259/dmfr.20150326. Epub 2016 Feb 19. Dentomaxillofac Radiol. 2016. PMID: 26892946 Free PMC article.
-
Diagnostic accuracy of digitized conventional radiographs by camera and scanner in detection of proximal caries.J Dent Res Dent Clin Dent Prospects. 2009 Fall;3(4):126-31. doi: 10.5681/joddd.2009.031. Epub 2009 Dec 15. J Dent Res Dent Clin Dent Prospects. 2009. PMID: 23230500 Free PMC article.
-
A comparison of the diagnostic accuracy of in vivo and in vitro photostimulable phosphor digital images in the detection of occlusal caries lesions.Dentomaxillofac Radiol. 2010 Jan;39(1):17-22. doi: 10.1259/dmfr/91657756. Dentomaxillofac Radiol. 2010. PMID: 20089739 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical