Uniform image quality achieved by tube current modulation using SD of attenuation in coronary CT angiography
- PMID: 17579170
- DOI: 10.2214/AJR.06.1201
Uniform image quality achieved by tube current modulation using SD of attenuation in coronary CT angiography
Erratum in
- AJR Am J Roentgenol. 2007 Aug;189(2):270
Abstract
Objective: The objective of our study was to evaluate whether the SD of CT attenuation values obtained from unenhanced scans of the left atrium is a reliable parameter for the individual modulation of tube current to achieve uniform image quality in coronary CT angiography (CTA).
Materials and methods: One hundred patients (59 men and 41 women) who were suspected to have coronary artery disease underwent coronary CTA using a 64-MDCT scanner. In addition to clinical studies, we also performed measurements on water phantoms. Tube current was modulated by the SD of the CT attenuation values measured from the left atrium on unenhanced images scanned at 300 mA. A modulation table was created from data obtained from the studies of water phantoms scanned at various tube currents. Other scanning parameters were identical to those used to obtain unenhanced and contrast-enhanced studies of the 100 patients. The SD values were measured from images scanned at an adjusted tube current, and the images of normal coronary and internal mammary arteries were graded. Radiation doses measured using the volume CT dose index (CTDI(vol)) were compared between the SD of the CT attenuation values and the modulation parameters suggested by the manufacturer of the scanner.
Results: Image quality was rated as grade 3 (low mottle) on a 4-grade scale by four observers for 92-94 of the 100 patients (average, 92.5%). The mean SD value at an adjusted tube current was 12.1 H with an SD of 0.758 H (target SD = 12 H). A radiation dose reduction of 9-45% was achieved in patients grouped by weight who weighed less than 70 kg, and a reduction of up to 71% was seen in individual cases.
Conclusion: Modulating tube current using the SD of CT attenuation values from the left atrium is a highly reliable method of achieving uniform image quality in coronary CTA.
Similar articles
-
Does a combination of dose modulation with fast gantry rotation time limit CT image quality?AJR Am J Roentgenol. 2008 Jul;191(1):140-4. doi: 10.2214/AJR.07.3019. AJR Am J Roentgenol. 2008. PMID: 18562737
-
Contrast injection protocols for coronary computed tomography angiography using a 64-detector scanner: comparison between patient weight-adjusted- and fixed iodine-dose protocols.Invest Radiol. 2008 Jul;43(7):512-9. doi: 10.1097/RLI.0b013e3181727505. Invest Radiol. 2008. PMID: 18580334 Clinical Trial.
-
Automated attenuation-based selection of tube voltage and tube current for coronary CT angiography: reduction of radiation exposure versus a BMI-based strategy with an expert investigator.J Cardiovasc Comput Tomogr. 2013 Sep-Oct;7(5):303-10. doi: 10.1016/j.jcct.2013.08.010. Epub 2013 Sep 26. J Cardiovasc Comput Tomogr. 2013. PMID: 24268117
-
Methods of plaque quantification and characterization by cardiac computed tomography.J Cardiovasc Comput Tomogr. 2009 Nov-Dec;3 Suppl 2:S91-8. doi: 10.1016/j.jcct.2009.10.012. Epub 2009 Oct 31. J Cardiovasc Comput Tomogr. 2009. PMID: 20129522 Review.
-
Cardiac computed tomography technology and dose-reduction strategies.Radiol Clin North Am. 2010 Jul;48(4):657-74. doi: 10.1016/j.rcl.2010.04.003. Epub 2010 Jun 11. Radiol Clin North Am. 2010. PMID: 20705164 Review.
Cited by
-
CT radiation dose optimization and estimation: an update for radiologists.Korean J Radiol. 2012 Jan-Feb;13(1):1-11. doi: 10.3348/kjr.2012.13.1.1. Epub 2011 Dec 23. Korean J Radiol. 2012. PMID: 22247630 Free PMC article. Review.
-
Coronary CT angiography with low radiation dose.Int J Cardiovasc Imaging. 2010 Feb;26 Suppl 1:17-25. doi: 10.1007/s10554-009-9576-5. Int J Cardiovasc Imaging. 2010. PMID: 20058080 Review.
-
Method for automatic tube current selection for obtaining a consistent image quality and dose optimization in a cardiac multidetector CT.Korean J Radiol. 2009 Nov-Dec;10(6):568-74. doi: 10.3348/kjr.2009.10.6.568. Korean J Radiol. 2009. PMID: 19885312 Free PMC article.
-
A practical approach for a patient-tailored dose protocol in coronary CT angiography using prospective ECG triggering.Int J Cardiovasc Imaging. 2016 Mar;32(3):531-8. doi: 10.1007/s10554-015-0802-z. Epub 2015 Nov 12. Int J Cardiovasc Imaging. 2016. PMID: 26563107 Free PMC article.
-
Image noise-based dose adaptation in dynamic volume CT of the heart: dose and image quality optimisation in comparison with BMI-based dose adaptation.Eur Radiol. 2014 Jan;24(1):86-94. doi: 10.1007/s00330-013-2980-1. Epub 2013 Aug 15. Eur Radiol. 2014. PMID: 23949725
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical