Reproducibility of DIAGNOdent 2095 and DIAGNOdent Pen measurements: results from an in vitro study on occlusal sites
- PMID: 17587296
- DOI: 10.1111/j.1600-0722.2007.00441.x
Reproducibility of DIAGNOdent 2095 and DIAGNOdent Pen measurements: results from an in vitro study on occlusal sites
Abstract
The aim of this in vitro study was to investigate the intra- and interdevice reproducibility of the DIAGNOdent 2095 (DD2095) and the new DIAGNOdent Pen (DDPen) on non-cavitated occlusal caries sites. Altogether, 241 sites on 90 molars were inspected twice with both devices and by four examiners. The intra/interdevice reproducibilities were assessed by means of the intraclass correlation coefficient (ICC) as well as by the range of the 95% limits of agreement of Bland & Altman. Additionally, a linear mixed-effects model was fitted with fixed effects for the visual scores and the devices used and random effects for occlusal sites and examiners within each measurement site. In result, the following values were determined for the intradevice reproducibility: DD2095: ICC 0.89, range 42.3; and DDPen: ICC 0.88, range 49.3. The interdevice reproducibility was found to be in the same order of magnitude (ICC 0.82, range 53.7). The comparison between both devices indicated significantly lower mean values for the DDPen. The estimates of the linear model show that there is a high variation between the sites that cannot be explained by the fixed effects. In conclusion, both devices showed an imperfect reproducibility, which indicate the usage as adjunct tool only in clinical practice.
Similar articles
-
The intra/inter-examiner reproducibility of the new DIAGNOdent Pen on occlusal sites.J Dent. 2007 Jun;35(6):509-12. doi: 10.1016/j.jdent.2007.02.001. Epub 2007 Mar 28. J Dent. 2007. PMID: 17395355
-
An in vitro study of the reliability of DIAGNOdent measurements.J Oral Rehabil. 2004 Sep;31(9):895-9. doi: 10.1111/j.1365-2842.2004.01315.x. J Oral Rehabil. 2004. PMID: 15369472
-
In vitro quantification of smooth surface caries with DIAGNOdent and the DIAGNOdent pen.Acta Odontol Scand. 2007 Feb;65(1):60-3. doi: 10.1080/00016350601058051. Acta Odontol Scand. 2007. PMID: 17354096
-
Comparison of two laser fluorescence devices for the detection of occlusal caries in vivo.Eur J Oral Sci. 2007 Aug;115(4):252-6. doi: 10.1111/j.1600-0722.2007.00456.x. Eur J Oral Sci. 2007. PMID: 17697163
-
[Diagnosis of occlusal caries lesions using laser fluorescence measurements].Ned Tijdschr Tandheelkd. 2002 Jan;109(1):3-7. Ned Tijdschr Tandheelkd. 2002. PMID: 11933610 Dutch.
Cited by
-
Treatment of various degrees of white spot lesions using resin infiltration-in vitro study.Prog Orthod. 2018 Aug 6;19(1):27. doi: 10.1186/s40510-018-0223-3. Prog Orthod. 2018. PMID: 30079435 Free PMC article.
-
Fluorescence-based methods for detecting caries lesions: systematic review, meta-analysis and sources of heterogeneity.PLoS One. 2013 Apr 4;8(4):e60421. doi: 10.1371/journal.pone.0060421. Print 2013. PLoS One. 2013. PMID: 23593215 Free PMC article.
-
Development, methodology and potential of the new Universal Visual Scoring System (UniViSS) for caries detection and diagnosis.Int J Environ Res Public Health. 2009 Sep;6(9):2500-9. doi: 10.3390/ijerph6092500. Epub 2009 Sep 23. Int J Environ Res Public Health. 2009. PMID: 19826559 Free PMC article.
-
Assessment of a new infrared laser transillumination technology (808 nm) for the detection of occlusal caries-an in vitro study.Lasers Med Sci. 2015 Sep;30(7):1873-9. doi: 10.1007/s10103-014-1704-3. Epub 2014 Dec 31. Lasers Med Sci. 2015. PMID: 25549960
-
Comparing the reliability of a new fluorescence camera with conventional laser fluorescence devices in detecting caries lesions in occlusal and smooth surfaces of primary teeth.Lasers Med Sci. 2011 Mar;26(2):157-62. doi: 10.1007/s10103-010-0757-1. Epub 2010 Feb 16. Lasers Med Sci. 2011. PMID: 20157753
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical