Temporal thermometry fails to track body core temperature during heat stress
- PMID: 17596768
- DOI: 10.1249/mss.0b013e318050ca3e
Temporal thermometry fails to track body core temperature during heat stress
Abstract
Purpose: The aim of this study was to assess the accuracy of temporal scanning thermometry in monitoring internal temperature increases during passive heating.
Methods: Sixteen subjects (5 males and 11 females) underwent a whole-body passive heat stress (water-perfused suit) to increase internal temperature. Temperatures were obtained with a temporal scanner and with an ingestible-pill telemetry system that tracks intestinal temperature. Temperatures were recorded while subjects were normothermic (34 degrees C water-perfusing suit) and every 10 min during passive heating (48 degrees C water-perfusing suit).
Results: Heart rate (ECG), mean skin temperature (weighted six-site average), skin blood flow (laser Doppler flowmetry), and sweat rate (capacitance hygrometry) were all significantly elevated at the end of heating (all P < 0.001). Pre-heat stress temporal-derived temperature was not different from intestinal temperature (36.98 +/- 0.09 vs 37.01 +/- 0.09 degrees C, respectively, P = 0.76). However, after 30 min of heating (the greatest duration of heating completed by all subjects), temporal-derived temperature decreased to below the pre-heat stress baseline (-0.22 +/- 0.11), whereas intestinal temperature increased by 0.39 +/- 0.07 degrees C (P < 0.001 between the two methods). After 50 min of heating (N = 11), intestinal-derived internal temperature increased by 0.70 +/- 0.09 degrees C, whereas temporal-derived temperature decreased by 0.29 +/- 0.10 degrees C (P < 0.001). The group average (+/- SEM) R2 and slope between the two methods were 0.29 +/- 0.08 and -0.34 +/- 0.14, respectively.
Conclusion: These results demonstrate that temporal scanning does not track internal temperature, as measured via intestinal temperature, during passive heating. Given these findings, it is recommended that this technique not be used to assess temperature in hyperthermic diaphoretic subjects.
Similar articles
-
The effects of face cooling on the prolactin response and subjective comfort during moderate passive heating in humans.Exp Physiol. 2006 Nov;91(6):1007-14. doi: 10.1113/expphysiol.2006.034629. Epub 2006 Aug 17. Exp Physiol. 2006. PMID: 16916892
-
The prolactin responses to active and passive heating in man.Exp Physiol. 2005 Nov;90(6):909-17. doi: 10.1113/expphysiol.2005.031294. Epub 2005 Sep 12. Exp Physiol. 2005. PMID: 16157657
-
Whole-body heating decreases skin vascular response to low orthostatic stress in the lower extremities.J Physiol Sci. 2006 Apr;56(2):157-64. doi: 10.2170/physiolsci.RP000406. Epub 2006 Apr 12. J Physiol Sci. 2006. PMID: 16839449
-
Accuracy of noninvasive core temperature measurement in acutely ill adults: the state of the science.Biol Res Nurs. 2006 Jul;8(1):24-34. doi: 10.1177/1099800406289151. Biol Res Nurs. 2006. PMID: 16766626 Review.
-
Body temperature variability (Part 1): a review of the history of body temperature and its variability due to site selection, biological rhythms, fitness, and aging.Altern Med Rev. 2006 Dec;11(4):278-93. Altern Med Rev. 2006. PMID: 17176167 Review.
Cited by
-
Validation of an ingestible temperature data logging and telemetry system during exercise in the heat.Temperature (Austin). 2016 Apr 6;3(2):208-219. doi: 10.1080/23328940.2016.1171281. eCollection 2016 Apr-Jun. Temperature (Austin). 2016. PMID: 27857951 Free PMC article.
-
Cooling intervention studies among outdoor occupational groups: A review of the literature.Am J Ind Med. 2020 Nov;63(11):988-1007. doi: 10.1002/ajim.23175. Epub 2020 Sep 4. Am J Ind Med. 2020. PMID: 32886396 Free PMC article.
-
Meta-analysis of heat-induced changes in cardiac function from over 400 laboratory-based heat exposure studies.Nat Commun. 2025 Mar 14;16(1):2543. doi: 10.1038/s41467-025-57868-6. Nat Commun. 2025. PMID: 40087302 Free PMC article.
-
Effects of Casein Hydrolysate Ingestion on Thermoregulatory Responses in Healthy Adults during Exercise in Heated Conditions: A Randomized Crossover Trial.Nutrients. 2020 Mar 24;12(3):867. doi: 10.3390/nu12030867. Nutrients. 2020. PMID: 32213899 Free PMC article. Clinical Trial.
-
Core-temperature sensor ingestion timing and measurement variability.J Athl Train. 2010 Nov-Dec;45(6):594-600. doi: 10.4085/1062-6050-45.6.594. J Athl Train. 2010. PMID: 21062183 Free PMC article. Clinical Trial.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical