Laboratory methods to improve SELDI peak detection and quantitation

Affiliations

Laboratory methods to improve SELDI peak detection and quantitation

Dominique Rollin et al. Proteome Sci. .

Abstract

Background: Protein profiling with surface-enhanced laser desorption-ionisation time-of-flight mass spectrometry (SELDI-TOF MS) is a promising approach for biomarker discovery. Some candidate biomarkers have been identified using SELDI-TOF, but validation of these can be challenging because of technical parameters that effect reproducibility. Here we describe steps to improve the reproducibility of peak detection.

Methods: SELDI-TOF mass spectrometry was performed using a system manufactured by Ciphergen Biosystems along with their ProteinChip System. Serum from 10 donors was pooled and used for all experiments. Serum was fractionated with Expression Difference Mapping kit-Serum Fractionation from the same company and applied to three different ProteinChips. The fractionations were run over a one month period to examine the contribution of sample batch and time to peak detection variability. Spectra were processed and peaks detected using the Ciphergen Express software and variance measured.

Results: Experimental parameters specific to the serum fraction and ProteinChip, including spot protocols (laser intensity and detector sensitivity) were optimized to decrease peak detection variance. Optimal instrument settings, regular calibration along with controlled sample handling and processing nearly doubled the number of peaks detected and decreased intensity variance.

Conclusion: This report assesses the variation across fractionated sera processed over a one-month period. The optimizations reported decreased the variance and increased the number of peaks detected.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Outline of the experimental workflow for SELDI-TOF optimization. QC serum was fractionated and run as 6 replicates of fractions F1, F3, F4, and F6 on IMAC, CM10 LS and H50 ProteinChips; as 6 replicates of F3 and F4 on CM10 HS and 6 replicates of F5 on CM10 LS ProteinChips.

Similar articles

Cited by

References

    1. Omenn GS, States DJ, Adamski M, Blackwell TW, Menon R, Hermjakob H, Apweiler R, Haab BB, Simpson RJ, Eddes JS, Kapp EA, Moritz RL, Chan DW, Rai AJ, Admon A, Aebersold R, Eng J, Hancock WS, Hefta SA, Meyer H, Paik YK, Yoo JS, Ping P, Pounds J, Adkins J, Qian X, Wang R, Wasinger V, Wu CY, Zhao X, Zeng R, Archakov A, Tsugita A, Beer I, Pandey A, Pisano M, Andrews P, Tammen H, Speicher DW, Hanash SM. Overview of the HUPO Plasma Proteome Project: results from the pilot phase with 35 collaborating laboratories and multiple analytical groups, generating a core dataset of 3020 proteins and a publicly-available database. Proteomics. 2005;5:3226–3245. doi: 10.1002/pmic.200500358. - DOI - PubMed
    1. Rai AJ, Stemmer PM, Zhang Z, Adam BL, Morgan WT, Caffrey RE, Podust VN, Patel M, Lim LY, Shipulina NV, Chan DW, Semmes OJ, Leung HC. Analysis of Human Proteome Organization Plasma Proteome Project (HUPO PPP) reference specimens using surface enhanced laser desorption/ionization-time of flight (SELDI-TOF) mass spectrometry: multi-institution correlation of spectra and identification of biomarkers. Proteomics. 2005;5:3467–3474. doi: 10.1002/pmic.200401320. - DOI - PubMed
    1. Banks RE, Stanley AJ, Cairns DA, Barrett JH, Clarke P, Thompson D, Selby PJ. Influences of Blood Sample Processing on Low-Molecular-Weight Proteome Identified by Surface-Enhanced Laser Desorption/Ionization Mass Spectrometry. Clin Chem. 2005;51:1637–1649. doi: 10.1373/clinchem.2005.051417. - DOI - PubMed
    1. Albrethsen J, Bogebo R, Olsen J, Raskov H, Gammeltoft S. Preanalytical and analytical variation of surface-enhanced laser desorption-ionization time-of-flight mass spectrometry of human serum. Clin Chem Lab Med. 2006;44:1243–1252. doi: 10.1515/CCLM.2006.228. - DOI - PubMed
    1. Liggett WS, Barker PE, Semmes OJ, Cazares LH. Measurement reproducibility in the early stages of biomarker development. Dis Markers. 2004;20:295–307. - PMC - PubMed

LinkOut - more resources