Flash electroretinography: normative values with surface skin electrodes and no pupil dilation using a standard stimulation protocol
- PMID: 17610098
- DOI: 10.1007/s10633-007-9065-x
Flash electroretinography: normative values with surface skin electrodes and no pupil dilation using a standard stimulation protocol
Abstract
Objectives: To demonstrate the clinical feasibility of using surface electrodes for recording flash electroretinography (ERG), using the stimulation standards of the International Society for the Clinical Electrophysiology of Vision (ISCEV), without pupillary dilation.
Methods: Eighteen physiologically normal volunteers were tested. Surface recording electrodes were placed over the middle third of the lower eyelid of each eye, and the reference electrode placed on the temple. The recording filters and method of stimulation for the standard flash cone, 30 Hz flicker, rod, standard combined and oscillatory responses were those recommended by the ISCEV. Four patients with a history of visual dysfunction were also examined.
Results: The amplitudes of the responses were noted to have a high signal to noise ratio, allowing us to average only a few responses. The waveform configurations were almost identical to those published using other recording methods. Patients with retinitis pigmentosa and early onset retinal dystrophy had abnormal responses, whereas patients with Stargardt's Syndrome and spasm of accommodation had responses within normal limits.
Conclusions: This study has established the clinical feasibility of performing flash ERG using surface skin electrodes, with no pupillary dilation and using the stimulation standards of the ISCEV.
Similar articles
-
Comparing DTL microfiber and Neuroline skin electrode in the Mini Ganzfeld ERG.BMC Ophthalmol. 2016 Aug 5;16:137. doi: 10.1186/s12886-016-0311-4. BMC Ophthalmol. 2016. PMID: 27491453 Free PMC article. Clinical Trial.
-
Evaluation of light- and dark-adapted ERGs using a mydriasis-free, portable system: clinical classifications and normative data.Doc Ophthalmol. 2018 Dec;137(3):169-181. doi: 10.1007/s10633-018-9660-z. Epub 2018 Oct 24. Doc Ophthalmol. 2018. PMID: 30357588
-
Accuracy and results of photopic flash electroretinogram performed with skin electrodes in infants.Eur J Ophthalmol. 2012 May-Jun;22(3):441-9. doi: 10.5301/ejo.5000021. Eur J Ophthalmol. 2012. PMID: 21748726
-
ISCEV standard for clinical multifocal electroretinography (mfERG) (2011 edition).Doc Ophthalmol. 2012 Feb;124(1):1-13. doi: 10.1007/s10633-011-9296-8. Epub 2011 Oct 30. Doc Ophthalmol. 2012. PMID: 22038576 Free PMC article. Review.
-
Electroretinography.Adv Exp Med Biol. 2018;1085:17-20. doi: 10.1007/978-3-319-95046-4_5. Adv Exp Med Biol. 2018. PMID: 30578478 Review.
Cited by
-
Use of multifocal electroretinograms to determine stage of glaucoma.PLoS One. 2023 Jan 12;18(1):e0278234. doi: 10.1371/journal.pone.0278234. eCollection 2023. PLoS One. 2023. PMID: 36634040 Free PMC article.
-
Comparison of DTL and gold cup skin electrodes for recordings of the multifocal electroretinogram.Doc Ophthalmol. 2023 Feb;146(1):67-78. doi: 10.1007/s10633-022-09912-9. Epub 2022 Dec 20. Doc Ophthalmol. 2023. PMID: 36536110 Free PMC article.
-
Enhanced S-cone function with preserved rod function: a new clinical phenotype.Mol Vis. 2011;17:2241-7. Epub 2011 Aug 18. Mol Vis. 2011. PMID: 21897746 Free PMC article.
-
A novel method to reduce noise in electroretinography using skin electrodes: a study of noise level, inter-session variability, and reproducibility.Int Ophthalmol. 2017 Apr;37(2):317-324. doi: 10.1007/s10792-016-0240-5. Epub 2016 Jun 9. Int Ophthalmol. 2017. PMID: 27278187 Free PMC article.
-
Brain trauma impacts retinal processing: photoreceptor pathway interactions in traumatic light sensitivity.Doc Ophthalmol. 2022 Jun;144(3):179-190. doi: 10.1007/s10633-022-09871-1. Epub 2022 Apr 20. Doc Ophthalmol. 2022. PMID: 35445376 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Molecular Biology Databases