Automatic detection of ineffective triggering and double triggering during mechanical ventilation
- PMID: 17611736
- DOI: 10.1007/s00134-007-0767-z
Automatic detection of ineffective triggering and double triggering during mechanical ventilation
Abstract
Objective: Imperfect patient-ventilator interaction is common during assisted ventilation, and the detection of clinically relevant mismatching requires visual monitoring of the ventilator screen. We have assessed the feasibility, sensitivity and specificity of an algorithm embedded in a ventilator system that is able to automatically detect the occurrence of ineffective triggering and double triggering in real time.
Design: Prospective study.
Setting: Respiratory intensive care unit.
Methods: Twenty patients undergoing pressure-support ventilation, either non-invasively (NIV, n=10) or conventionally ventilated (n=10), were studied.
Measurements: The detection of ineffective triggering and double triggering from the algorithm was compared by two operators with the "real" occurrence of the phenomena as assessed using the transdiaphragmatic pressure (Pdi).
Results: Seven of the 20 patients exhibited gross mismatching, while in the remaining patients patient-ventilator mismatching was artificially induced using a pressure control, with a low respiratory rate. Ineffective triggering and double triggering were identified by the operators in 507 and 19 of the 3343 analyzed breaths, respectively. False positives were significantly more frequent in the NIV group than with conventional ventilation. The algorithm had an overall sensitivity of 91% and specificity of 97%. Specificity was statistically higher in the conventional ventilated group than with NIV (99% vs. 95%, p<0.05).
Conclusions: We have demonstrated the feasibility and efficacy of a new algorithm to detect the occurrence of impaired patient-ventilator interaction during mechanical ventilation in real time. This software may help the clinician in the identification of this problem, which has been shown to have important clinical consequences.
Similar articles
-
Patient-ventilator asynchrony during noninvasive ventilation: a bench and clinical study.Chest. 2012 Aug;142(2):367-376. doi: 10.1378/chest.11-2279. Chest. 2012. PMID: 22406958 Clinical Trial.
-
Detecting ineffective triggering in the expiratory phase in mechanically ventilated patients based on airway flow and pressure deflection: feasibility of using a computer algorithm.Crit Care Med. 2008 Feb;36(2):455-61. doi: 10.1097/01.CCM.0000299734.34469.D9. Crit Care Med. 2008. PMID: 18091543 Clinical Trial.
-
Effects of the flow waveform method of triggering and cycling on patient-ventilator interaction during pressure support.Intensive Care Med. 2003 Nov;29(11):1950-9. doi: 10.1007/s00134-003-1703-5. Epub 2003 Mar 29. Intensive Care Med. 2003. PMID: 12665996 Clinical Trial.
-
Equipment needs for noninvasive mechanical ventilation.Eur Respir J. 2002 Oct;20(4):1029-36. doi: 10.1183/09031936.02.00404202. Eur Respir J. 2002. PMID: 12412700 Review.
-
Ventilator modes and settings during non-invasive ventilation: effects on respiratory events and implications for their identification.Thorax. 2011 Feb;66(2):170-8. doi: 10.1136/thx.2010.142661. Epub 2010 Oct 14. Thorax. 2011. PMID: 20947891 Review.
Cited by
-
Automatic detection of AutoPEEP during controlled mechanical ventilation.Biomed Eng Online. 2012 Jun 20;11:32. doi: 10.1186/1475-925X-11-32. Biomed Eng Online. 2012. PMID: 22715924 Free PMC article.
-
Patient-Ventilator Synchronization During Non-invasive Ventilation: A Pilot Study of an Automated Analysis System.Front Med Technol. 2021 Jul 7;3:690442. doi: 10.3389/fmedt.2021.690442. eCollection 2021. Front Med Technol. 2021. PMID: 35047935 Free PMC article.
-
Automatic detection of patient-ventilator asynchrony by spectral analysis of airway flow.Crit Care. 2011 Jul 12;15(4):R167. doi: 10.1186/cc10309. Crit Care. 2011. PMID: 21749683 Free PMC article.
-
Timing of inspiratory muscle activity detected from airway pressure and flow during pressure support ventilation: the waveform method.Crit Care. 2022 Jan 30;26(1):32. doi: 10.1186/s13054-022-03895-4. Crit Care. 2022. PMID: 35094707 Free PMC article.
-
The oesophageal balloon for respiratory monitoring in ventilated patients: updated clinical review and practical aspects.Eur Respir Rev. 2023 May 17;32(168):220186. doi: 10.1183/16000617.0186-2022. Print 2023 Jun 30. Eur Respir Rev. 2023. PMID: 37197768 Free PMC article. Review.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources